FDA Adverse Event Injury Summary report: N

CHARGING SYSTEM

MDR report key: 2971072 · Received February 15, 2013

Report

Report Number
1627487-2013-04198
Event Type
Injury
Date Received
February 15, 2013
Date of Event
January 25, 2013
Report Date
January 25, 2013
Manufacturer
ST. JUDE MEDICAL - NEUROMODULATION
Product Code
LGW
PMA / PMN Number
P010032
Removal / Correction Number
1627487-07262012-001-C
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THIS CHARGER MODEL WAS ASSOCIATED WITH A FIELD CORRECTION. MANUFACTURER'S EVALUATION: CORRECTIVE AND PREVENTIVE ACTION (CAPA) INVESTIGATION WAS PERFORMED. EVALUATION CODES: RESULTS: POCKET HEATING WAS CONFIRMED. THE INVESTIGATION FOR (B)(4) ASSOCIATED WITH HEATING WHILE CHARGING (POCKET HEATING) CONCLUDED THAT THE CHARGER WAS CAPABLE OF TRANSFERRING ENERGY TO THE IPG AT A RATE THAT WOULD CAUSE HEATING OF THE IPG AND/OR CHARGING WAND OF SUFFICIENT ELEVATED TEMPERATURE TO CAUSE PAIN AND BURNS. THE HEATING WHILE CHARGING WAS DETERMINED TO BE EXACERBATED BY OFF-AXIS CHARGING OF SHALLOW IMPLANTED IPGS AND THAT ALL CHARGERS WERE CAPABLE OF ELEVATED HEATING. SJM HAS LIMITED INFORMATION RELATED TO THE PATIENT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PATIENT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PATIENT'S PHYSICIAN REGARDING MEDICAL HISTORY.

Description of Event or Problem · 1

DEVICE 2 OF 2. REFERENCE MFR REPORT: 1627487-2013-04197. IT WAS REPORTED THE PATIENT HAD DIFFICULTY CHARGING, AND WAS EXPERIENCING HEATING WHILE CHARGING THE IPG. THE PATIENT ALSO REPORTED THE AC ADAPTER WOULD GET HOT. THE SJM REPRESENTATIVE MET WITH THE PATIENT, AND WAS ABLE TO ESTABLISH COMMUNICATION WITHOUT INCIDENT. DURING THE CHARGING SESSION, THE PATIENT STATED SHE COULD NOT DISTINGUISH IF THERE WAS HEATING. A REPLACEMENT CHARGING SYSTEM WAS SENT TO THE PATIENT. FOLLOW UP IDENTIFIED THE ISSUE HAD RESOLVED. IT WAS REPORTED THE HEATING ISSUE WAS ONLY PRESENT IF THE PATIENT WAS MOVING AROUND WHILE CHARGING. THE PATIENT WAS ADVISED TO CHARGE MORE FREQUENTLY FOR LESS TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
69638 CHARGING SYSTEM SCS CHARGING SYSTEM LGW ST. JUDE MEDICAL - NEUROMODULATION 3711 3779187

Patients

Seq Age Sex Outcome Treatment
1 54 YR Required Intervention IMPLANT:| SCS LEAD: MODEL 3166| SCS LEAD: MODEL 3163| IMPLANT:| SCS ANCHOR: MODEL 1192 (2)| IMPLANT:| SCS EXTENSION: MODEL 3341 (2)| IMPLANT: