FDA Adverse Event
Injury
Summary report: N
S-ROM*STM STD,30 NK,16X11X150
MDR report key: 2970959
·
Received February 21, 2013
Report
- Report Number
- 1818910-2013-03982
- Event Type
- Injury
- Date Received
- February 21, 2013
- Date of Event
- January 31, 2013
- Report Date
- January 31, 2013
- Manufacturer
- DEPUY IRELAND
- Product Code
- JDI
- PMA / PMN Number
- K851422
- Removal / Correction Number
- Z-1749/1816-2011
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE ASR PLATFORM WAS VOLUNTARILY RECALLED FROM THE MARKET IN AUGUST 2010, AND THE ASR PRODUCT CODES ARE NOW CONSIDERED INACTIVE. FURTHER INVESTIGATION OF THIS INDIVIDUAL INCIDENT WILL NOT BE UNDERTAKEN, AS THERE IS AN ONGOING INVESTIGATION REGARDING THE ROOT CAUSE(S) AND/OR CORRECTIVE ACTIONS. (B)(4). DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD THE PRODUCT AND/OR ADDITIONAL INFORMATION BE RECEIVED, THE INVESTIGATION WILL BE RE-OPENED.
Description of Event or Problem · 1
PATIENT WAS REVISED DUE TO ELEVATED COCR LEVELS, ALVAL, CYST WAS PRESENT, ALSO NOTED INTEROPERATIVELY WAS CORROSION AT THE NECK OF THE STEM AND INSIDE THE SLEEVE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 76696 | S-ROM*STM STD,30 NK,16X11X150 | STEM | JDI | DEPUY IRELAND | 1222161 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 53 YR | Required Intervention |