FDA Adverse Event Injury Summary report: N

ASR UNI FEMORAL IMPL SIZE 45

MDR report key: 2970957 · Received February 21, 2013

Report

Report Number
1818910-2013-03979
Event Type
Injury
Date Received
February 21, 2013
Date of Event
January 31, 2013
Report Date
June 27, 2013
Manufacturer
DEPUY INTERNATIONAL
Product Code
KWA
PMA / PMN Number
K040627
Removal / Correction Number
Z-1749/1816-2011
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE ASR PLATFORM WAS VOLUNTARILY RECALLED FROM THE MARKET IN AUGUST 2010, AND THE ASR PRODUCT CODES ARE NOW CONSIDERED INACTIVE. FURTHER INVESTIGATION OF THIS INDIVIDUAL INCIDENT WILL NOT BE UNDERTAKEN, AS THERE IS AN ONGOING INVESTIGATION REGARDING THE ROOT CAUSE(S) AND/OR CORRECTIVE ACTIONS. (B)(4). DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD THE PRODUCT AND/OR ADDITIONAL INFORMATION BE RECEIVED, THE INVESTIGATION WILL BE RE-OPENED.

Additional Manufacturer Narrative · 1

DEPUY STILL CONSIDERS THIS COMPLAINT CLOSED.

Description of Event or Problem · 1

PATIENT WAS REVISED DUE TO ELEVATED COCR LEVELS, ALVAL, CYST WAS PRESENT, ALSO NOTED INTEROPERATIVELY WAS CORROSION AT THE NECK OF THE STEM AND INSIDE THE SLEEVE.

Description of Event or Problem · 1

**UPDATE** (B)(4) 2013-LITIGATION PAPERS RECEIVED ALLEGING THAT THE PATIENT SUFFERS FROM PAIN. THERE IS NO NEW ADDITIONAL INFORMATION THAT WOULD CHANGE THE INVESTIGATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
76247 ASR UNI FEMORAL IMPL SIZE 45 FEMORAL HEAD KWA DEPUY INTERNATIONAL 2434790

Patients

Seq Age Sex Outcome Treatment
1 53 YR Required Intervention