FDA Adverse Event Malfunction Summary report: N

PK DISSECTING FORCEPS INSTRUMENT

MDR report key: 2970814 · Received February 20, 2013

Report

Report Number
2955842-2013-00574
Event Type
Malfunction
Date Received
February 20, 2013
Date of Event
January 9, 2013
Report Date
January 25, 2013
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. ENGINEERING WAS UNABLE TO CONFIRM THE REPORTED FAILURE MODE. THE GRIPS HAD A VERY SMALL OFFSET AT THE TIPS MEASURING ROUGHLY .005. THIS OFFSET IS BELOW THE MAXIMUM ALLOWABLE SPECIFICATION OF .010, THEREFORE THE INSTRUMENT WAS ACCEPTABLE AND OK TO USE. ENGINEERING EVALUATION ALSO FOUND THAT THE PITCH UP CABLE WAS BROKEN AT THE DISTAL CLEVIS HUB. THE CABLE SEGMENT THAT CONTAINS THE CRIMP REMAINED INSTALLED IN THE CLEVIS. THE CLEVIS DID NOT EXHIBIT ANY DAMAGE OR WEAR MARKS. THE OTHER CABLES AT WRIST WERE NOT DAMAGED. THE DISTAL END OF THE MAIN TUBE HAD VARIOUS SCRATCH MARKS WITH LIGHT MATERIAL REMOVAL. THE SCRATCHES WERE .070 - .215 IN LENGTH AND WERE NOT ALIGNED WITH THE TUBE AXIS. ENGINEERING CONCLUDED THAT THE DAMAGE TO THE TUBE WAS LIKELY DUE TO MISHANDLING/MISUSE. NO OTHER DAMAGE WAS FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER; THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT PRIOR TO STARTING A DA VINCI SI SURGICAL PROCEDURE, THE TIPS ON THE PK DISSECTING FORCEPS INSTRUMENT WERE OBSERVED TO BE MISALIGNED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
74865 PK DISSECTING FORCEPS INSTRUMENT ENDOSCOPIC ELECTROSURGICAL INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420227-03 M10120531 286

Patients

Seq Age Sex Outcome Treatment
1 DA VINCI SI SURGICAL SYSTEM, ACCESSORIES & ESU.