FDA Adverse Event Malfunction Summary report: N

PK DISSECTING FORCEPS INSTRUMENT

MDR report key: 2970801 · Received February 20, 2013

Report

Report Number
2955842-2013-00573
Event Type
Malfunction
Date Received
February 20, 2013
Date of Event
January 9, 2013
Report Date
January 25, 2013
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING OBSERVED THAT ONE GRIP WAS BENT, CAUSING SIDE TO SIDE MISALIGNMENT OF GRIPS. THERE IS A .055 OFFSET AT THE TIPS, INDICATING THAT OVERLOADING AT THE TIP OCCURRED. THE INSTRUMENT PASSED ELECTRICAL CONTINUITY TESTING. ENGINEERING EVALUATION ALSO FOUND THAT THE PITCH CABLE WAS BROKEN AT THE DISTAL CLEVIS HUB. THE CABLE SEGMENT THAT CONTAINS THE CRIMP REMAINED INSTALLED IN THE CLEVIS. THE CLEVIS DID NOT EXHIBIT ANY DAMAGE OR WEAR MARKS.

Description of Event or Problem · 1

IT WAS REPORTED THAT PRIOR TO STARTING A DA VINCI SI SURGICAL PROCEDURE, THE TIPS ON THE PK DISSECTING FORCEPS INSTRUMENT WERE OBSERVED TO BE MISALIGNED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
74975 PK DISSECTING FORCEPS INSTRUMENT ENDOSCOPIC ELECTROSURGICAL INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420227-03 M10120523 553

Patients

Seq Age Sex Outcome Treatment
1 DA VINCI SI SURG SYS. INSTRUMENTS AND ACCESSORIES