FDA Adverse Event
Malfunction
Summary report: N
PK DISSECTING FORCEPS INSTRUMENT
MDR report key: 2970801
·
Received February 20, 2013
Report
- Report Number
- 2955842-2013-00573
- Event Type
- Malfunction
- Date Received
- February 20, 2013
- Date of Event
- January 9, 2013
- Report Date
- January 25, 2013
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K050369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING OBSERVED THAT ONE GRIP WAS BENT, CAUSING SIDE TO SIDE MISALIGNMENT OF GRIPS. THERE IS A .055 OFFSET AT THE TIPS, INDICATING THAT OVERLOADING AT THE TIP OCCURRED. THE INSTRUMENT PASSED ELECTRICAL CONTINUITY TESTING. ENGINEERING EVALUATION ALSO FOUND THAT THE PITCH CABLE WAS BROKEN AT THE DISTAL CLEVIS HUB. THE CABLE SEGMENT THAT CONTAINS THE CRIMP REMAINED INSTALLED IN THE CLEVIS. THE CLEVIS DID NOT EXHIBIT ANY DAMAGE OR WEAR MARKS.
Description of Event or Problem · 1
IT WAS REPORTED THAT PRIOR TO STARTING A DA VINCI SI SURGICAL PROCEDURE, THE TIPS ON THE PK DISSECTING FORCEPS INSTRUMENT WERE OBSERVED TO BE MISALIGNED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 74975 | PK DISSECTING FORCEPS INSTRUMENT | ENDOSCOPIC ELECTROSURGICAL INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420227-03 | M10120523 553 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | DA VINCI SI SURG SYS. INSTRUMENTS AND ACCESSORIES |