PENUMBRA COIL 400
Report
- Report Number
- 3005168196-2013-00067
- Event Type
- Malfunction
- Date Received
- February 20, 2013
- Date of Event
- January 21, 2013
- Report Date
- January 21, 2013
- Manufacturer
- PENUMBRA, INC.
- Product Code
- HCG
- PMA / PMN Number
- K103305
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- ME, US
- Reporter Occupation
- PHYSICIAN
Narratives
CONCLUSION: THIS DEVICE IS AVAILABLE FOR RETURN AND A FOLLOW-UP MDR WILL BE SUBMITTED UPON COMPLETION OF THE DEVICE INVESTIGATION.
RESULTS: THE COIL IS STUCK INSIDE OF THE ORANGE SHEATH THAT IT GETS DEPLOYED THROUGH. THE PUSHER ASSEMBLY STILL HAS THE PET LOCK IN TACT AT THE PROXIMAL END. THIS PUSHER ASSEMBLY ALSO HAS THE CONSTRAINT BALL INSIDE THE DISTAL DETACHMENT TIP (DDT) WITH (0.8 CM) OF SR-WIRE ATTACHED. THE SR WIRE IS BROKEN. THE COIL IS NOT ATTACHED TO THE PUSHER ASSEMBLY. CONCLUSION: THE COMPLAINT HAS BEEN EVALUATED AND CONFIRMED. THE DESCRIPTION STATES THAT DURING RETRIEVAL OF THE COIL, THE COIL DETACHED INSIDE THE SHEATH. AFTER INVESTIGATING THE RETURNED PRODUCT, IT APPEARS THAT DURING RETRIEVAL OF THE COIL IT BECAME STUCK INSIDE THE SHEATH AND WHEN THE PHYSICIAN CONTINUED TO PULL THE COIL BACK, THE TENSILE STRENGTH OF THE SR WIRE MATERIAL WAS EXCEEDED CAUSING THE WIRE TO BREAK AND THE COIL TO DETACH. THE MANUFACTURING RECORDS FOR THIS LOT WERE REVIEWED AND DID NOT REVEAL ANY OUTSTANDING DISCREPANCIES, DESIGN, OR QUALITY CONCERNS.
PATIENT WAS BEING TREATED FOR TWO INTRACRANIAL ANEURYSMS. THE LESION INVOLVED IN THIS INCIDENT WAS A CAROTID TERMINUS ANEURYSM LOCATED ON THE PATIENT'S RIGHT SIDE. IT MEASURED 12MM BY 11MM BY 11MM. THE PHYSICIAN DELIVERED THREE PC400 COILS SUCCESSFULLY INTO THE ANEURYSM USING A PXSLIMSTR CATHETER WHICH WAS PLACED THROUGH A NEURON MAX 088 90CM LONG SHEATH. THE PHYSICIAN TRIED TO DELIVER A 3X10 CURVE EXTRA SOFT, BUT WAS KICKED OUT OF THE ANEURYSM. HE RESHEATHED SUCCESSFULLY. HE THEN TRIED A SECOND TIME BUT THE COIL AGAIN COULD NOT BE PLACED. HE THEN TRIED TO RESHEATH THIS COIL. HE SUCCESSFULLY GOT THE COIL OUT OF THE SLIM MICROCATHETER, BUT THE COIL DETACHED INSIDE THE COIL INTRODUCER. HE THEN PROCEEDED TO COMPLETE THE CASE WITH TWO ADDITIONAL PC400 COILS. THERE WAS NO IMPACT ON THE PATIENT RELATED TO THIS INCIDENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 74641 | PENUMBRA COIL 400 | HCG, KRD | HCG | PENUMBRA, INC. | F27776 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 61 YR |