FDA Adverse Event Malfunction Summary report: N

PK DISSECTING FORCEPS

MDR report key: 2970574 · Received February 20, 2013

Report

Report Number
2955842-2013-00566
Event Type
Malfunction
Date Received
February 20, 2013
Date of Event
December 20, 2012
Report Date
January 22, 2013
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K061260
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. ENGINEERING OBSERVED A BROKEN CABLE. THE PITCH CABLE IS BROKEN AT THE DISTAL CLEVIS HUB. THE CABLE SEGMENT THAT CONTAINS THE CRIMP HAS STILL REMAINED IN THE CLEVIS. THE CLEVIS DOES NOT EXHIBIT EXCESSIVE DAMAGE. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER, THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DA VINCI SI SURGICAL PROCEDURE, THE PK DISSECTING FORCEPS INSTRUMENT WOULD NOT OPEN AND/OR CLOSE. AFTER THE INSTRUMENT WAS PULLED OUT, IT WAS DISCOVERED THAT A CABLE WAS BROKEN. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
75028 PK DISSECTING FORCEPS ENDOSCOPIC ELECTROSURGICAL INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420227-03 M10121106 179

Patients

Seq Age Sex Outcome Treatment
1 DA VINCI SI SYSTEM INSTRUMENTS AND ACCESSORIES