FDA Adverse Event
Malfunction
Summary report: N
PK DISSECTING FORCEPS
MDR report key: 2970574
·
Received February 20, 2013
Report
- Report Number
- 2955842-2013-00566
- Event Type
- Malfunction
- Date Received
- February 20, 2013
- Date of Event
- December 20, 2012
- Report Date
- January 22, 2013
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K061260
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE INSTRUMENT WAS RETURNED AND EVALUATED. ENGINEERING OBSERVED A BROKEN CABLE. THE PITCH CABLE IS BROKEN AT THE DISTAL CLEVIS HUB. THE CABLE SEGMENT THAT CONTAINS THE CRIMP HAS STILL REMAINED IN THE CLEVIS. THE CLEVIS DOES NOT EXHIBIT EXCESSIVE DAMAGE. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER, THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING A DA VINCI SI SURGICAL PROCEDURE, THE PK DISSECTING FORCEPS INSTRUMENT WOULD NOT OPEN AND/OR CLOSE. AFTER THE INSTRUMENT WAS PULLED OUT, IT WAS DISCOVERED THAT A CABLE WAS BROKEN. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 75028 | PK DISSECTING FORCEPS | ENDOSCOPIC ELECTROSURGICAL INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420227-03 | M10121106 179 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | DA VINCI SI SYSTEM INSTRUMENTS AND ACCESSORIES |