FENESTRATED BIPOLAR FORCEPS INSTRUMENT
Report
- Report Number
- 2955842-2013-00555
- Event Type
- Injury
- Date Received
- February 20, 2013
- Date of Event
- January 21, 2013
- Report Date
- January 21, 2013
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K050369
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AR
- Reporter Occupation
- NURSE
Narratives
THE INSTRUMENT WAS RETURNED AND EVALUATED. ENGINEERING WAS UNABLE TO CONFIRM THE CUSTOMER REPORTED FAILURE MODE. THE INSTRUMENT PASSED RECOGNITION AND ENGAGEMENT TESTING WHEN TESTED ON AN ISI IN-HOUSE IS3000 SYSTEM. THE INSTRUMENT MOVED INTUITIVELY WITH A FULL RANGE OF MOTION IN ALL DIRECTIONS, THE GRIPS OPENED AND CLOSED PROPERLY AND PERFORMED CAUTERY FUNCTION WITH NO TROUBLE FOUND. ON (B)(6) 2013, ISI CONTACTED THE ISI CLINICAL SALES REPRESENTATIVE THAT WAS PRESENT DURING THE SURGICAL PROCEDURE. THE CSR INDICATED THAT APPROXIMATELY 30 - 40 MINUTES INTO THE SURGICAL PROCEDURE, WHILE THE SURGEON WAS GRASPING TISSUE THAT WAS BEING DISSECTED FOR REMOVAL, THE SURGEON OBSERVED ENERGY COMING FROM THE FENESTRATED BIPOLAR FORCEPS INSTRUMENT WITHOUT ANY ACTIVATION OF CAUTERY ENERGY FROM THE FOOT PEDAL ON THE SURGEON SIDE CART (SSC). THE CSR INDICATED THAT THE SITE PERFORMED TROUBLESHOOTING OF THE ISSUE AND WHEN THE SURGEON ATTEMPTED TO GRASP TISSUE A 2ND TIME, THE ISSUE RECURRED. THE CSR INDICATED ENERGY ACTIVATED A 3RD TIME FROM THE INSTRUMENT WHEN THE SURGEON DEMONSTRATED TO HIM WHAT HAD OCCURRED DURING TROUBLESHOOTING. THE CSR INDICATED THAT THE FENESTRATED BIPOLAR FORCEPS INSTRUMENT WAS IMMEDIATELY REPLACED AND THE ISSUE WAS RESOLVED. THE CSR INDICATED THAT A VALLEY LAB FORCE TRIAD ELECTRICAL SURGICAL UNIT (ESU) AND A VALLEY LAB DISPOSABLE ESU CABLE WERE USED IN CONJUNCTION WITH THE FENESTRATED BIPOLAR FORCEPS INSTRUMENT. THE CSR INDICATED THAT IT IS UNKNOWN IF THE SITE INSPECTED THE ESU PRIOR TO USE; HOWEVER, THE SITE DID INSPECT THE ESU DURING TROUBLESHOOTING AND FOUND THAT THE ESU CABLE WAS INSTALLED PROPERLY. THE CSR INDICATED THAT THE SETTINGS ON THE SSC WERE CORRECT AND THAT THE SURGICAL PROCEDURE WAS NOT RECORDED. THE CSR INDICATED THAT THE SURGICAL PROCEDURE WAS COMPLETED AND THERE WAS NO IMPACT TO THE PATIENT, EXCEPT TO THE TISSUE BEING REMOVED AS PART OF THE SURGICAL PROCEDURE. THE CSR ALSO INDICATED THAT THERE WAS NO REPORT BY THE SURGEON THAT THE PATIENT EXPERIENCED ANY POST-SURGICAL COMPLICATIONS AS A RESULT OF THE REPORTED EVENT. ON (B)(6) 2013, THE SITE'S DA VINCI SI SURGICAL SYSTEM WAS EVALUATED BY AN ISI FIELD SERVICE ENGINEER. THE FSE WAS UNABLE TO REPLICATE THE FAILURE MODE EXPERIENCED BY THE SITE AS INSPECTION OF THE SITE'S SURGICAL SYSTEM FOUND THAT IT FUNCTIONED WITHIN SPECIFICATIONS. AS OF (B)(6) 2013, THERE HAVE BEEN NO REPORTED RECURRENCES OF THE ISSUE AT THIS HOSPITAL.
IT WAS REPORTED THAT DURING A DA VINCI SI HYSTERECTOMY PROCEDURE, THE FENESTRATED BIPOLAR FORCEPS INSTRUMENT ACTIVATED WITHOUT THE SURGEON ACTIVATING CAUTERY ENERGY FROM THE FOOT PEDAL ON THE SURGEON SIDE CART. NO PATIENT HARM WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 74904 | FENESTRATED BIPOLAR FORCEPS INSTRUMENT | ENDOSCOPIC ELECTROSURGICAL INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420205-05 | M10120613 768 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other | DA VINCI SI SURG. SYSTEM, ACCESSORIES AND AN ESU. |