FDA Adverse Event
Injury
Summary report: N
PERMOBIL C300 PS
MDR report key: 2969699
·
Received February 11, 2013
Report
- Report Number
- 1221084-2013-00002
- Event Type
- Injury
- Date Received
- February 11, 2013
- Date of Event
- December 31, 2012
- Report Date
- February 8, 2013
- Manufacturer
- PERMOBIL, INC.
- Product Code
- ITI
- PMA / PMN Number
- K041219
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
Description of Event or Problem · 1
THE BACK CANES ON THE BACKREST OF THE POWER WHEELCHAIR FAILED AND THE CLIENT FELL BACKWARD. THE CLIENT SUFFERED A CUT ON HIS HEAD. DEALER PROVIDED INFORMATION INDICATING THAT PUBLIC TRANSPORTATION INAPPROPRIATELY SECURED THE WHEELCHAIR BY ATTACHING THE TIE DOWNS TO THE BACK CANES. THIS SITUATION PLACED SIGNIFICANT STRESS ON THE BACK CANES CAUSING THEM TO BREAK. THE WHEELCHAIR HAS DESIGNATED AREAS PROPERLY MARKED WHERE TIE DOWNS SHOULD BE ATTACHED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 60049 | PERMOBIL C300 PS | POWER WHEELCHAIR | ITI | PERMOBIL, INC. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |