VENTED AUTOFEED HUMIDIFICATION CHAMBER
Report
- Report Number
- 9611451-2013-00114
- Event Type
- Malfunction
- Date Received
- February 18, 2013
- Date of Event
- January 21, 2013
- Report Date
- January 24, 2013
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LIMITED
- Product Code
- BTT
- PMA / PMN Number
- K934140
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE COMPLAINT MR290V VENTED AUTOFEED HUMIDIFICATION CHAMBER IS EN ROUTE TO FISHER & PAYKEL HEALTHCARE FOR INVESTIGATION. WE WILL PROVIDE A FOLLOW-UP REPORT ONCE WE HAVE COMPLETED OUR INVESTIGATION.
(B)(4). METHOD: THE COMPLAINT MR290V VENTED AUTOFEED HUMIDIFICATION CHAMBER WAS RETURNED TO FISHER & PAYKEL HEALTHCARE (FPH) IN (B)(4) AND WAS VISUALLY INSPECTED. RESULTS: VISUAL INSPECTION REVEALED THAT THERE WAS A BREAK IN THE FEEDSET TUBING AT THE CONNECTION TO THE CHAMBER DOME. THE SURFACE OF THE BREAK WAS ROUGH (NOT SMOOTHLY CUT). A LOT CHECK REVEALED ONE OTHER COMPLAINTS OF THIS NATURE FOR LOT NUMBER 120829. CONCLUSION: THE DAMAGE APPEARED TO BE CAUSED BY THE TUBE BEING PULLED AWAY FROM THE CHAMBER, POSSIBLY DUE TO THE FEEDSET BEING CAUGHT OR UNDER TENSION. ALL CHAMBERS ARE PRESSURE TESTED BEFORE THEY LEAVE THE PRODUCTION LINE, AND ANY HOLES, LEAKS, OR BREAKS IN THE FEEDSET ARE IDENTIFIED DURING THIS PROCESS. THIS SUGGESTS THAT THE DAMAGE OCCURRED POST PRODUCTION. THE USER INSTRUCTIONS WHICH ACCOMPANY THE MR290 CHAMBER STATE THE FOLLOWING: "SET APPROPRIATE VENTILATOR ALARM." "PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT." (B)(4).
A HOSPITAL IN (B)(6) REPORTED VIA A DISTRIBUTOR THAT AN MR290V VENTED AUTOFEED HUMIDIFICATION CHAMBER HAD A DAMAGED WATER FEEDSET TUBE. THIS WAS OBSERVED DURING USE ON A PATIENT. NO PATIENT CONSEQUENCE WAS REPORTED.
A HOSPITAL IN (B)(6) REPORTED VIA A DISTRIBUTOR THAT AN MR290V VENTED AUTOFEED HUMIDIFICATION CHAMBER HAD A DAMAGED WATER FEEDSET TUBE. THIS WAS OBSERVED DURING USE ON A PATIENT. NO PATIENT CONSEQUENCE WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 70811 | VENTED AUTOFEED HUMIDIFICATION CHAMBER | BTT | BTT | FISHER & PAYKEL HEALTHCARE LIMITED | MR290V | 120829 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |