FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 2965167 · Received February 15, 2013

Report

Report Number
3004209178-2013-91100
Event Type
Malfunction
Date Received
February 15, 2013
Date of Event
December 3, 2012
Report Date
February 4, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP WAS RECEIVED WITH A MOTOR ERROR ALARM DURING THE BASIC OCCLUSION TEST, AND THE DEVICE ALARMED DURING THE PRIME TEST AS A RESULT OF A LOOSE DRIVE SUPPORT DISK.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE INSULIN PUMP ALARMED MOTOR ERROR. THE BLOOD GLUCOSE READING WAS 173MG/DL. TROUBLESHOOTING WAS PERFORMED. THE DEVICE WAS NOT EXPOSED TO A HIGH MAGNETIC FIELD. ADVISED THE CALLER THAT THE INSULIN PUMP WOULD BE REPLACED AND REVERT TO BACK UP PLAN. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
66849 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-722NAS

Patients

Seq Age Sex Outcome Treatment
1 53 YR