PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 2965167
·
Received February 15, 2013
Report
- Report Number
- 3004209178-2013-91100
- Event Type
- Malfunction
- Date Received
- February 15, 2013
- Date of Event
- December 3, 2012
- Report Date
- February 4, 2013
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP WAS RECEIVED WITH A MOTOR ERROR ALARM DURING THE BASIC OCCLUSION TEST, AND THE DEVICE ALARMED DURING THE PRIME TEST AS A RESULT OF A LOOSE DRIVE SUPPORT DISK.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE INSULIN PUMP ALARMED MOTOR ERROR. THE BLOOD GLUCOSE READING WAS 173MG/DL. TROUBLESHOOTING WAS PERFORMED. THE DEVICE WAS NOT EXPOSED TO A HIGH MAGNETIC FIELD. ADVISED THE CALLER THAT THE INSULIN PUMP WOULD BE REPLACED AND REVERT TO BACK UP PLAN. NO FURTHER INFORMATION WAS PROVIDED.