FDA Adverse Event Malfunction Summary report: N

ELECTRODE, PACEMAKER, PERMANENT

MDR report key: 2965154 · Received February 15, 2013

Report

Report Number
2649622-2013-01743
Event Type
Malfunction
Date Received
February 15, 2013
Date of Event
December 28, 2012
Report Date
February 27, 2013
Manufacturer
MPRI
Product Code
DTB
PMA / PMN Number
P930039
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT HAD A SUSPECTED RIGHT VENTRICULAR (RV) PERFORATION AND PERICARDIAL EFFUSION. THE RV LEAD WAS REMOVED. DURING THE IMPLANT ATTEMPT OF THE REPLACEMENT RV LEAD, THERE WAS UNRESOLVABLE SENSING AND NOISE. A DIFFERENT LEAD WAS IMPLANTED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
68723 ELECTRODE, PACEMAKER, PERMANENT DTB MPRI 5076-45

Patients

Seq Age Sex Outcome Treatment
1 00050 YR Hospitalization| L| R