PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 2965149
·
Received February 15, 2013
Report
- Report Number
- 3004209178-2013-91104
- Event Type
- Malfunction
- Date Received
- February 15, 2013
- Date of Event
- December 30, 2012
- Report Date
- February 6, 2013
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP ALARMED DURING THE BASIC OCCLUSION TEST, AND THE DEVICE WAS UNABLE TO PRIME DURING THE PRIME TEST AS A RESULT OF A LOOSE DRIVE SUPPORT DISK. THE INSULIN PUMP HAD MISSING END CAP STICKER.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE CUSTOMER REPORTED HAD ISSUES DURING PRIME/REWIND. THE BLOOD GLUCOSE READING WAS 9.3MMOL/L. THE CALLER STATED THAT THE INSULIN PUMP ALARMED. ADVISED THE CALLER THAT THE INSULIN PUMP WOULD BE REPLACED. NO FURTHER INFORMATION WAS PROVIDED.