FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO

MDR report key: 2965129 · Received February 15, 2013

Report

Report Number
2182208-2013-00403
Event Type
Injury
Date Received
February 15, 2013
Date of Event
December 25, 2012
Report Date
December 31, 2012
Manufacturer
RICE CREEK MFG
Product Code
LWS
PMA / PMN Number
P920015
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TU
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. (B)(4).

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE PROXIMAL SEGMENT OF THE LEAD WAS RETURNED, ANALYZED. ANALYSIS PERFORMED SHOWED NO ANOMALIES WERE FOUND, THE OUTER INSULATION OF THE LEAD WAS EXTRINSICALLY BREACHED DUE TO A CUT. IT WAS VISUALLY NOTED THAT THE LEAD INDICATED APPARENT EXPLANT DAMAGE. PERFORMANCE DATA COLLECTED FROM THE DEVICE INDICATED THE IMPEDANCE TREND ON THE RV (RIGHT VENTRICULAR) PACING LEAD WAS RISING, AND OVERSENSING DUE TO NON-PHYSIOLOGIC SIGNALS/SIC (SENSING INTEGRITY COUNTER).

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT WAS ADMITTED TO THE HOSPITAL WITH UNWANTED SHOCKS. ON INTERROGATION THE RIGHT VENTRICULAR (RV) LEAD WAS NOTED TO HAVE HIGH IMPEDANCE. THE LEAD WAS EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
68031 SPRINT QUATTRO DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS RICE CREEK MFG 6944

Patients

Seq Age Sex Outcome Treatment
1 00059 YR Hospitalization| L| R