FDA Adverse Event Injury Summary report: N

CONCERTO II CRT-D

MDR report key: 2965076 · Received February 15, 2013

Report

Report Number
3004209178-2013-02302
Event Type
Injury
Date Received
February 15, 2013
Report Date
December 17, 2012
Manufacturer
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
Product Code
NIK
PMA / PMN Number
P010031
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. ANALYSIS DID NOT DETECT ANY PERFORMANCE ISSUES, BUT THE CALCULATED LONGEVITY RESULT OF 86% IS LESS THAN THE PREDICTED VALUE BASED ON THE AVAILABLE PROGRAMMED PARAMETERS. IN THE ABSENCE OF THE COMPLETE PROGRAMMING HISTORY, IT IS NOT POSSIBLE TO DETERMINE WHY THE LONGEVITY DID NOT MATCH THE PREDICTED MODEL.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCTS: 1788TC COMPETITOR PACING LEAD (B)(6) 2007; 7000 COMPETITOR DEFIB LEAD (B)(6) 2007. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE MAY HAVE REACHED THE ELECTIVE REPLACEMENT INDICATOR (ERI) PREMATURELY. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
67352 CONCERTO II CRT-D DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO NIK MEDTRONIC MED REL MEDTRONIC PUERTO RICO D274TRK

Patients

Seq Age Sex Outcome Treatment
1 00054 YR Hospitalization| R 1058T COMPETITOR PACING LEAD