VIRTUOSO DR
Report
- Report Number
- 9614453-2013-00446
- Event Type
- Injury
- Date Received
- February 15, 2013
- Report Date
- November 23, 2012
- Manufacturer
- IPG MFG SWITZERLAND
- Product Code
- LWS
- PMA / PMN Number
- P980016
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BE
- Reporter Occupation
- PHYSICIAN
Narratives
THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. THIS MODEL NUMBER IS NOT APPROVED FOR DISTRIBUTION IN THE UNITED STATES, HOWEVER, IT IS SIMILAR TO A DEVICE MARKETED IN THE U.S. THE EVENT IS BEING REPORTED DUE TO AN ALLEGED MALFUNCTION. (B)(4).
IT WAS REPORTED THAT THE PATIENT'S IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) HAD SHORTENED BATTERY LIFE DUE TO THE HOLTER POSITION BEING "ON". THE ICD WAS EXPLANTED AND REPLACED. THE PHYSICIAN WAS VERY ANGRY OF NOT BEING MADE AWARE THAT THE HOLTER BEING IN THE "ON" POSITION WOULD CAUSE EARLY BATTERY DEPLETION. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 67206 | VIRTUOSO DR | LWS | IPG MFG SWITZERLAND | D164AWG |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization| R |