FDA Adverse Event Malfunction Summary report: N

PULSE-GENERATOR, PACEMAKER, IMPLANTABLE

MDR report key: 2965065 · Received February 15, 2013

Report

Report Number
2182208-2013-00424
Event Type
Malfunction
Date Received
February 15, 2013
Date of Event
November 14, 2012
Report Date
November 14, 2012
Manufacturer
RICE CREEK MFG
Product Code
DXY
PMA / PMN Number
P890003
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. (B)(4).

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: ANALYSIS CONFIRMED THE REPORTED EVENT; A PRINTED CIRCUIT BOARD WAS FOUND OUT OF ELECTRICAL SPECIFICATION.

Description of Event or Problem · 1

IT WAS REPORTED THERE WAS ECG (ELECTROCARDIOGRAM) NOISE. THE PROGRAMMER WAS RETURNED FOR REPAIR. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
67797 PULSE-GENERATOR, PACEMAKER, IMPLANTABLE DXY RICE CREEK MFG 2090

Patients

Seq Age Sex Outcome Treatment
1