DRUG ELUTING PERMANENT RIGHT VENTRICULAR (RV) OR RIGHT ATRIAL (RA) PACEMAKER ELE
Report
- Report Number
- 2649622-2013-01924
- Event Type
- Malfunction
- Date Received
- February 15, 2013
- Date of Event
- November 28, 2012
- Report Date
- November 28, 2012
- Manufacturer
- MPRI
- Product Code
- NVN
- PMA / PMN Number
- P090013
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- PHYSICIAN
Narratives
PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED, ANALYZED, AND NO ANOMALIES WERE FOUND.
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).
CORRECTED INFORMATION: NO EVAL EXPLAIN CODE. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IT WAS REPORTED THAT DURING THE IMPLANT PROCEDURE, THE LEAD HAD NOISE. THE NOISE PERSISTED AFTER CHANGING THE CABLES AND ANALYZER. THE LEAD WAS NOT USED, AND ANOTHER LEAD WAS USED TO COMPLETE THE PROCEDURE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
IT WAS LATER REPORTED THAT THE PROGRAMMER WAS PLUGGED INTO AN EXTENSION CABLE THAT WAS NOT GROUNDED PROPERLY.