CONSULTA CRT-D
Report
- Report Number
- 3004209178-2013-02126
- Event Type
- Injury
- Date Received
- February 15, 2013
- Report Date
- December 7, 2012
- Manufacturer
- MEDTRONIC MED REL MEDTRONIC PUERTO RICO
- Product Code
- NIK
- PMA / PMN Number
- P010031
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- PHYSICIAN
Narratives
EVALUATION SUMMARY: (B)(4) THE DEVICE WAS RETURNED, ANALYZED AND THE DEVICE EXPERIENCED NORMAL BATTERY DEPLETION.
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCTS: 4076 IMPLANTABLE PACING LEAD (B)(6) 2009; 4196 IMPLANTABLE PACING LEAD (B)(6) 2009. (B)(4).
IT WAS REPORTED THAT THE DEVICE HAD A SHORTER THAN ANTICIPATED BATTERY LONGEVITY DUE TO HIGH LEFT VENTRICULAR OUTPUT. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
IT WAS REPORTED THAT THE DEVICE HAD A SHORTER THAN ANTICIPATED BATTERY LONGEVITY. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 68147 | CONSULTA CRT-D | DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO | NIK | MEDTRONIC MED REL MEDTRONIC PUERTO RICO | D224TRK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00076 YR | Hospitalization| R | 6947 IMPLANTABLE DEFIB LEAD |