FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO SECURE S

MDR report key: 2964946 · Received February 15, 2013

Report

Report Number
2649622-2013-01536
Event Type
Injury
Date Received
February 15, 2013
Report Date
December 5, 2012
Manufacturer
MPRI
Product Code
LWS
PMA / PMN Number
P920015
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE FULL LEAD IN SEGMENTS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. ALSO, PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS RECEIVED, ANALYZED AND NO ANOMALIES WERE FOUND.

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. (B)(4).

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT EXPERIENCED PERICARDIAL EFFUSION, POSSIBLY BECAUSE THE HELIX WAS SCREWED TOO DEEP IN THE TISSUE. THE LEAD WAS EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
66774 SPRINT QUATTRO SECURE S DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MPRI 693565

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| L| R (B)(4) IMPLANTABLE CRT-D