CAPSUREFIX
Report
- Report Number
- 2649622-2013-01926
- Event Type
- Injury
- Date Received
- February 15, 2013
- Report Date
- December 14, 2012
- Manufacturer
- MPRI
- Product Code
- NVN
- PMA / PMN Number
- P930039
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. (B)(4).
PRODUCT EVENT SUMMARY: THE PROXIMAL SEGMENT OF THE LEAD WAS RETURNED AND ANALYZED. ANALYSIS WAS PERFORMED AND NO ANOMALIES WERE FOUND. THERE WERE NO ANOMALIES OBSERVED REGARDING THE RETURNED LEAD 3 CM PROXIMAL SEGMENT. THE FULL LEAD WAS NOT RETURNED.
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IT WAS REPORTED THAT THE ATRIAL LEAD HAD A POLARITY SWITCH AND HAD DECREASED IMPEDANCE. THE LEAD WAS CAPPED AND WAS REPLACED WITH A NEW LEAD. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT
IT WAS ADDITIONALLY REPORTED THAT THE LEAD WAS EXPLANTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 68502 | CAPSUREFIX | DRUG ELUTING PERMANENT RIGHT VENTRICULAR (RV) OR RIGHT ATRIAL (RA) PACEMAKER ELE | NVN | MPRI | 4068 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization| R | (B)(4) IMPLANTABLE PULSE GENERATOR (IPG) |