FDA Adverse Event Injury Summary report: N

ATTAIN OTW

MDR report key: 2964860 · Received February 15, 2013

Report

Report Number
2649622-2013-01942
Event Type
Injury
Date Received
February 15, 2013
Date of Event
December 31, 2012
Report Date
December 31, 2012
Manufacturer
MPRI
Product Code
LWP
PMA / PMN Number
P010015
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCTS: 5076 IMPLANTABLE PACING LEAD (B)(6) 2010; 6947 IMPLANTABLE TACHY LEAD, (B)(6) 2010. (B)(4).

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THERE WAS EARLY BATTERY DEPLETION, ALONG WITH CHRONIC HIGH LV (LEFT VENTRICULAR) THRESHOLD AND POSSIBLE LEAD DI SLODGEMENT. THE DEVICE AND LV LEAD WERE EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
67925 ATTAIN OTW PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE LWP MPRI 419488

Patients

Seq Age Sex Outcome Treatment
1 00055 YR Hospitalization| R (B)(4) IMPLANTABLE CRT-D