FDA Adverse Event Injury Summary report: N

ELECTRODE, PACEMAKER, PERMANENT

MDR report key: 2964728 · Received February 15, 2013

Report

Report Number
2649622-2013-02127
Event Type
Injury
Date Received
February 15, 2013
Date of Event
December 18, 2012
Report Date
March 7, 2013
Manufacturer
MPRI
Product Code
DTB
PMA / PMN Number
P930039
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED AND ANALYSIS FOUND THAT THE OUTER INSULATION WAS BREACHED WITH METAL ION OXIDATION (MIO). THERE WAS ALSO COSMETIC MIO.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4). CONCOMITANT PRODUCT: 4092 IMPLANTABLE PACING LEAD (B)(6) 2004.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE ATRIAL LEAD P WAVES AND IMPEDANCE HAD BEEN DECLINING OVER THE PAST FEW MONTHS. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
66675 ELECTRODE, PACEMAKER, PERMANENT DTB MPRI 4568-45

Patients

Seq Age Sex Outcome Treatment
1 00084 YR Hospitalization| R KDR701 IMPLANTABLE PULSE GENERATOR (IPG)