FDA Adverse Event Injury Summary report: N

CONCERTO II CRT-D

MDR report key: 2964720 · Received February 15, 2013

Report

Report Number
3004209178-2013-02319
Event Type
Injury
Date Received
February 15, 2013
Report Date
February 18, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO., JUNCOS
Product Code
NIK
PMA / PMN Number
P010031
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY #THE ACTUAL DEVICE WAS NOT RECEIVED FOR EVALUATION. WE DID RECEIVE PERFORMANCE DATA COLLECTED FROM THE DEVICE AND HAVE ANALYZED THEDATA. ANALYSIS INDICATED THAT ALERTS WERE TRIGGERED DUE TO LOW BATTERY VOLTAGE AND RECOMMENDED REPLACEMENT TIME (RRT).

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCT: 6949 DEFIB LEAD (B)(6) 2005; 5076 NON-DEFIB LEAD (B)(6) 2005. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE TRIGGERED THE ELECTIVE REPLACEMENT INDICATOR (ERI) AND PREMATURE BATTERY DEPLETION WAS SUSPECTED. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE TRIGGERED THE ELECTIVE REPLACEMENT INDICATOR (ERI) AND PREMATURE BATTERY DEPLETION WAS SUSPECTED. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
66427 CONCERTO II CRT-D DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO NIK MEDTRONIC PUERTO RICO OPERATIONS CO., JUNCOS D274TRK

Patients

Seq Age Sex Outcome Treatment
1 00084 YR Hospitalization| R 4194 NON-DEFIB LEAD