FDA Adverse Event Injury Summary report: N

MARQUIS

MDR report key: 2964715 · Received February 15, 2013

Report

Report Number
9614453-2013-00425
Event Type
Injury
Date Received
February 15, 2013
Report Date
February 19, 2013
Manufacturer
IPG MFG SWITZERLAND
Product Code
LWS
PMA / PMN Number
P980016
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

PRODUCT (B)(4) - THE ACTUAL DEVICE WAS NOT RECEIVED FOR EVALUATION. WE DID RECEIVE PERFORMANCE DATA COLLECTED FROM THE DEVICE AND HAVE ANALYZED THE DATA. THERE WAS ONE POWER ON RESET FOR WRITE TO LOCKED RAM, ADDR=0C77, DATA=89 ON (B)(4) 2012. THERE WAS ONE PATIENT ALERT FOR POWER ON REST ON (B)(4) 2012.

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE HAD AN ELECTRICAL RESET. THE DEVICE IS STILL IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED ASA RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
68689 MARQUIS DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS IPG MFG SWITZERLAND 7230CX

Patients

Seq Age Sex Outcome Treatment
1 00050 YR Required Intervention