FDA Adverse Event Injury Summary report: N

SECURA VR

MDR report key: 2964642 · Received February 15, 2013

Report

Report Number
9614453-2013-00411
Event Type
Injury
Date Received
February 15, 2013
Date of Event
June 4, 2012
Report Date
December 3, 2012
Manufacturer
MEDTRONIC S.A. (SMO)
Product Code
LWS
PMA / PMN Number
P980016
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. THIS MODEL NUMBER IS NOT APPROVED FOR DISTRIBUTION IN THE UNITED STATES, HOWEVER, IT IS SIMILAR TO A DEVICE MARKETED IN THE U.S. THE EVENT IS BEING REPORTED DUE TO AN ALLEGED MALFUNCTION. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT RECEIVED INAPPROPRIATE THERAPY DUE TO T WAVE OVERSENSING. THE DEVICE WAS REPROGRAMMED TO INCREASE THE SENSITIVITY THRESHOLD AND REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
69299 SECURA VR DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MEDTRONIC S.A. (SMO) D234VRC

Patients

Seq Age Sex Outcome Treatment
1 00048 YR Life Threatening| R