FDA Adverse Event Malfunction Summary report: N

ADJ PIN COLLET 2.0-3.2MM

MDR report key: 2964590 · Received February 15, 2013

Report

Report Number
0001811755-2013-00295
Event Type
Malfunction
Date Received
February 15, 2013
Date of Event
February 6, 2013
Report Date
February 6, 2013
Manufacturer
STRYKER INSTRUMENTS-KALAMAZOO
Product Code
KIJ
PMA / PMN Number
K972367
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE QUALITY SPECIALIST CONFIRMED THE REPORTED EVENT AND NOTED METAL SHAVINGS WERE FOUND AND THE DRIVESHAFT HAD SCORE MARKS. WEAR ON THE DRIVESHAFT CAN PRODUCE METAL SHAVINGS AND CAUSE THE DEVICE TO NOT GRIP A PIN. THEREFORE, IT IS LIKELY THE REPORTED EVENT WAS DUE TO DRIVESHAFT SCORING. THE DEVICE WILL NOT BE REPAIRED AND RETURNED TO THE CUSTOMER, THE DEVICE WAS DISCARDED BY THE MANUFACTURER.

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS IS IN PROGRESS; ADDITIONAL INFORMATION WILL BE SUBMITTED ONCE THE QUALITY INVESTIGATION IS COMPLETED.

Description of Event or Problem · 1

THE ADJ PIN COLLET 2-3.2 MM WAS SENT FOR SERVICE AND DURING FAILURE ANALYSIS IT WAS NOTED THAT THERE WERE METAL SHAVINGS PRESENT ON THE SHAFT OF THE DEVICE. THERE WAS NO PATIENT INVOLVEMENT AND NO ADVERSE CONSEQUENCES ASSOCIATED WITH THE DEVICE.

Description of Event or Problem · 1

THE ADJ PIN COLLET 2-3.2 MM WAS SENT FOR SERVICE AND DURING FAILURE ANALYSIS IT WAS NOTED THAT THERE WERE METAL SHAVINGS PRESENT ON THE SHAFT OF THE DEVICE. THERE WAS NO PATIENT INVOLVEMENT AND NO ADVERSE CONSEQUENCES ASSOCIATED WITH THE DEVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
69050 ADJ PIN COLLET 2.0-3.2MM INSTRUMENT, SURGICAL, ORTHOPEDIC, DC-POWERED MOTOR AND ACCESSORY/ATTACHMENT KIJ STRYKER INSTRUMENTS-KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1