FDA Adverse Event Malfunction Summary report: N

CORE IMPACTION DRILL

MDR report key: 2964585 · Received February 15, 2013

Report

Report Number
0001811755-2013-00301
Event Type
Malfunction
Date Received
February 15, 2013
Date of Event
January 31, 2013
Report Date
January 31, 2013
Manufacturer
STRYKER INSTRUMENTS-KALAMAZOO
Product Code
DZJ
PMA / PMN Number
K040369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DURING THE DEVICE EVALUATION, THE REPORTED OVERHEATING COULD NOT BE CONFIRMED. HOWEVER, A WORN COUPLER WAS FOUND, WHICH MAY LEAD TO INTERMITTENT FRICTION AND MAY CAUSE OVERHEATING.

Additional Manufacturer Narrative · 1

ADDITIONAL INFORMATION WILL BE SUBMITTED ONCE THE QUALITY INVESTIGATION HAS BEEN PERFORMED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING EQUIPMENT TESTING CONDUCTED AT THE USER FACILITY, THE DRILL WAS OVERHEATING. NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE ALLEGED WITH THIS EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING EQUIPMENT TESTING CONDUCTED AT THE USER FACILITY, THE DRILL WAS OVERHEATING. NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE ALLEGED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
67061 CORE IMPACTION DRILL DRIVER, WIRE, AND BONE DRILL, MANUAL DZJ STRYKER INSTRUMENTS-KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1