- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OR, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).
PRODUCT EVENT SUMMARY (B)(4) THE FULL LEAD WAS RETURNED AND ANALYZED. THE PRIMARY ANALYSIS FINDING NOTED NO ANOMALIES WERE FOUND. THE DISTAL ELECTRODE WAS COVERED WITH BODY TISSUE/FIBROTIC GROWTH AND WAS COVERED WITH BLOOD. VISUAL ANALYSIS NOTED THE LEAD HAD APPARENT EXPLANT DAMAGE.
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD PULLED BACK MONTHS AFTER IMPLANT. THE LEAD WAS DAMAGED DURING EXPLANT. THE LEAD WAS REPLACED WITH A NEW ONE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS THE RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 66656 | CAPSUREFIX NOVUS | ELECTRODE, PACEMAKER, PERMANENT | DTB | MPRI | 407652 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00061 YR | Hospitalization| R | (B)(4) IMPLANTABLE CARDIOVERTER DEFIBRILLATOR |