FDA Adverse Event Injury Summary report: N

CAPSUREFIX NOVUS

MDR report key: 2964543 · Received February 15, 2013

Report

Report Number
2649622-2013-02044
Event Type
Injury
Date Received
February 15, 2013
Date of Event
December 14, 2012
Report Date
March 6, 2013
Manufacturer
MPRI
Product Code
DTB
PMA / PMN Number
P930039
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OR, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY (B)(4) THE FULL LEAD WAS RETURNED AND ANALYZED. THE PRIMARY ANALYSIS FINDING NOTED NO ANOMALIES WERE FOUND. THE DISTAL ELECTRODE WAS COVERED WITH BODY TISSUE/FIBROTIC GROWTH AND WAS COVERED WITH BLOOD. VISUAL ANALYSIS NOTED THE LEAD HAD APPARENT EXPLANT DAMAGE.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD PULLED BACK MONTHS AFTER IMPLANT. THE LEAD WAS DAMAGED DURING EXPLANT. THE LEAD WAS REPLACED WITH A NEW ONE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS THE RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
66656 CAPSUREFIX NOVUS ELECTRODE, PACEMAKER, PERMANENT DTB MPRI 407652

Patients

Seq Age Sex Outcome Treatment
1 00061 YR Hospitalization| R (B)(4) IMPLANTABLE CARDIOVERTER DEFIBRILLATOR