NA
Report
- Report Number
- 0002090040-2013-00003
- Event Type
- Malfunction
- Date Received
- February 15, 2013
- Date of Event
- January 11, 2013
- Report Date
- January 21, 2013
- Manufacturer
- STRYKER SUSTAINABILITY SOLUTIONS LAKELAND
- Product Code
- NUJ
- PMA / PMN Number
- K111600
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
(B)(4).
THE COMPLAINT DEVICE WAS RETURNED TO STRYKER SUSTAINABILITY SOLUTIONS (SSS) FOR AN EVALUATION. A VISUAL EXAMINATION OF THE RETURNED DEVICE DID REVEAL ESCHAR BUILD-UP IN THE GUIDING SLOTS. THE JAWS WERE UNABLE TO OPEN SINCE THE BLADE WAS DISLOCATED/NOT FULLY RETRACTED. THE BLADE WAS REALIGNED WITH THE GUIDING SLOTS AND WAS ABLE TO RETRACT. ALL HANDLE MECHANICS BECAME FUNCTIONAL ONCE THE BLADE WAS REALIGNED. WHILE THE JAWS WERE IN THE LOCKED POSITION, THEY WERE INSPECTED AND FOUND TO BE SLIGHTLY MISALIGNED. THE BLADE TRIGGER WAS TESTED AND VERIFIED TO MAINTAIN MOVEMENT; HOWEVER, THERE WAS A SLIGHT STICKY FEEL WHEN ACTIVATED AS A RESULT OF THE MISALIGNED JAWS. BASED ON THE EVALUATION OF THE DEVICE, THIS TYPE OF FAILURE CAN BE CAUSED BY CLAMPING ON LARGE, RIGID TISSUE. THE DESIGN OF THE LF4200 ALLOWS FOR LARGER BITES OF TISSUE THAN OTHER DEVICES. HOWEVER, IT IS CRITICAL THAT THE USER DOES NOT PLACE TOO MUCH TISSUE IN THE JAWS DURING USE. IF THE JAWS ARE OVERFILLED, IT IS POSSIBLE FOR THE CUTTING MECHANISM TO BE DAMAGED, POTENTIALLY RESULTING IN DIFFICULTY OPENING THE JAWS OR INJURY RESULTING TO THE USER OR PATIENT. SSS INSTRUCTIONS FOR USE STATES: "KEEP THE INSTRUMENT JAWS CLEAN. BUILD-UP OF ESCHAR MAY REDUCE SEALING AND/OR CUTTING EFFECTIVENESS. WIPE JAW SURFACES AND EDGES WITH A WET GAUZE PAD AS NEEDED." "DO NOT USE THIS DEVICE ON VESSELS IN EXCESS OF 7MM IN DIAMETER." "TO ENSURE PROPER FUNCTION, ELIMINATE TENSION ON THE TISSUE WHILE SEALING AND CUTTING." "AVOID OVERFILLING THE INSTRUMENT JAWS WITH TISSUE. THIS MAY DAMAGE THE CUTTING MECHANISM OR CAUSE THE BLADE TO DEPLOY OUTSIDE OF ITS GUIDING FEATURES, POSSIBLY RESULTING IN DIFFICULTY OPENING THE JAWS OR UNINTENDED INJURY TO THE USER OR PATIENT." "VISUALLY CONFIRM THAT THE JAWS HAVE REACHED THE CLOSED POSITION PRIOR TO ACTIVATING THE CUTTER. FAILURE TO DO SO MAY DAMAGE THE CUTTING MECHANISM OR CAUSE THE BLADE TO DEPLOY OUTSIDE OF ITS GUIDING FEATURES, POSSIBLY RESULTING IN DIFFICULTY OPENING THE JAWS OR UNINTENDED INJURY TO THE USER OR PATIENT." "PRIOR TO ACTIVATING THE CUTTER, CONFIRM THAT THE JAWS ARE IN THE CLOSED POSITION. SPACING BETWEEN JAWS MUST BE LESS THAN TWO MILLIMETERS." THE DEVICE HISTORY RECORD FOR THE RETURNED DEVICE INDICATES THAT THE COMPLAINT DEVICE PASSED ALL APPLICABLE INSPECTIONS AND TESTS PRIOR TO RELEASE. THIS REPORT IS BEING FILED AS A MALFUNCTION DUE TO SSS BEING IN A 2 YEAR REPORTING CYCLE DUE TO MDR REPORT 1056128-2012-00042, WHERE THE DEVICE HAD TO BE CUT AWAY FROM THE PATIENT'S TISSUE WHEN IT BECAME STUCK EVEN THOUGH THERE WAS NO PATIENT INJURY IN THIS EVENT. THE REPORTED EVENT IS NOT OCCURRING MORE FREQUENTLY OR WITH GREATER SEVERITY THAN IS USUAL FOR THE DEVICE.
IT WAS REPORTED THAT "MIDWAY THROUGH THE CASE" THE LIGASURE "DEVICE WOULD NOT OPEN." NO PATIENT INJURY WAS REPORTED BY THE USER FACILITY.
IT WAS REPORTED THAT "MIDWAY THROUGH THE CASE" THE LIGASURE "DEVICE WOULD NOT OPEN." NO PATIENT INJURY WAS REPORTED BY THE USER FACILITY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 67493 | NA | NUJ | NUJ | STRYKER SUSTAINABILITY SOLUTIONS LAKELAND | LF4200 | 2089588 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |