CAPSUREFIX NOVUS MRI
Report
- Report Number
- 2649622-2013-02058
- Event Type
- Injury
- Date Received
- February 15, 2013
- Date of Event
- December 17, 2012
- Report Date
- December 20, 2012
- Manufacturer
- MPRI
- Product Code
- NVN
- PMA / PMN Number
- P090013
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCT: 5086MRI IMPLANTABLE PACING LEAD (B)(6) 2012. (B)(4).
PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THE LOW VOLTAGE DISTAL END ELECTRODE WAS COVERED WITH BLOOD.
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IT WAS REPORTED THAT THE PATIENT DEVELOPED PERICARDITIS WITHIN 24 HOURS POST IMPLANT. A MICRO-PERFORATION WAS SUSPECTED SO THE RIGHT VENTRICULAR AND RIGHT ATRIAL LEADS WERE BOTH EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 67952 | CAPSUREFIX NOVUS MRI | DRUG ELUTING PERMANENT RIGHT VENTRICULAR (RV) OR RIGHT ATRIAL (RA) PACEMAKER ELE | NVN | MPRI | 5086MRI58 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00023 YR | Hospitalization| L| R | RVDR01 IMPLANTABLE PULSE GENERATOR |