FDA Adverse Event Injury Summary report: N

PROTECTA XT DR

MDR report key: 2964489 · Received February 15, 2013

Report

Report Number
3004209178-2013-02293
Event Type
Injury
Date Received
February 15, 2013
Date of Event
December 21, 2012
Report Date
December 31, 2012
Manufacturer
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
Product Code
LWS
PMA / PMN Number
P980016
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCTS: 5076 IMPLANTABLE PACING LEAD, (B)(6) 2012. (B)(4).

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED, ANALYZED, AND NO ANOMALIES WERE FOUND.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THERE WAS NOISE ON THE RIGHT VENTRICULAR (RV) SENSING CIRCUIT AND THE IMPEDANCE WAS JUMPING TO HIGH LEVELS. THE SET SCREW WAS FOUND TO BE FINE AND THE LEAD APPEARED NORMAL THROUGH THE ANALYZER, SO AN UNKNOWN HEADER CONNECTION ISSUE WAS SUSPECTED. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
67295 PROTECTA XT DR DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MEDTRONIC MED REL MEDTRONIC PUERTO RICO D314DRM

Patients

Seq Age Sex Outcome Treatment
1 00068 YR Hospitalization| R 6935M IMPLANTABLE TACHY LEAD