CARPENTIER-EDWARDS PERIMOUNT RSR PERICARDIAL BIOPROSTHESIS
Report
- Report Number
- 2015691-2013-19300
- Event Type
- Injury
- Date Received
- February 15, 2013
- Date of Event
- December 20, 2012
- Report Date
- January 17, 2013
- Manufacturer
- EDWARDS LIFESCIENCES
- Product Code
- DYE
- PMA / PMN Number
- P860057/S001
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NV, US
- Reporter Occupation
- OTHER
Narratives
DEVICE NOT RETURNED. UNFORTUNATELY, THE EXPLANTED DEVICE WAS DISCARDED BY THE HOSPITAL; THEREFORE, THE SOURCE OF THE REPORTED CALCIFICATION COULD NOT BE ASSESSED. ALTHOUGH THE ROOT CAUSE OF THIS EVENT CANNOT BE CONFIRMED, IT APPEARS THAT THE STENOSIS WAS LIKELY CAUSED BY THE CALCIFICATION NOTED. CALCIFICATION IS A WELL RECOGNIZED FAILURE MODE OF BIOPROSTHETIC VALVES. THE MECHANISMS FOR BIOPROSTHETIC HEART VALVE TISSUE CALCIFICATION ARE NOT FULLY UNDERSTOOD. MANY FACTORS CAN CONTRIBUTE TO THE ONSET AND PROPAGATION OF CALCIFICATION INCLUDING PATIENT RELATED (E.G. PATIENT AGE, DISEASE STATE, IMMUNE STATUS, AND OTHER CO-MORBIDITIES), PHARMACOLOGICAL, AND INTRINSIC PROPERTIES OF THE VALVE ITSELF. IT IS WIDELY UNDERSTOOD THAT PATIENTS WITH CHRONIC RENAL DISEASE AND PRIOR HISTORY OF CALCIFIC STENOSIS OF THE NATIVE VALVE MAY BE PREDISPOSED TO BIOPROSTHETIC CALCIFICATION. IN THIS CASE, THE HEALTHCARE PROVIDER HAS INDICATED THAT THIS WAS AGE RELATED CALCIFICATION. NO FURTHER ACTIONS ARE POSSIBLE WITH THE AVAILABLE INFORMATION.
THROUGH FOLLOW UP, COPIES OF THE PATIENT'S MEDICAL RECORDS WERE RECEIVED WHICH SUPPORT THE INITIAL REPORT OF STENOSIS AND CALCIFICATION. PER THE OP REPORT, TEE SHOWED CRITICAL AORTIC STENOSIS WITH PANEL AORTIC VALVE NONFUNCTIONAL AORTIC VALVE PREVIOUSLY PLACED 10 YEARS AGO. THE VALVE WAS CONFIRMED TO BE SEVERELY STENOTIC UPON EXPLANT. THE DEVICE WAS REPLACED WITH A NEW EDWARDS BIOPROSTHETIC VALVE. ECHO SHOWED NO PERIVALVULAR LEAK. THE PATIENT TOLERATED THE PROCEDURE WELL.
EDWARDS LIFESCIENCES MAINTAINS AN IMPLANT PATIENT REGISTRY. THIS REGISTRY IS A PATIENT TRACKING MECHANISM FOR SERIALIZED EDWARDS IMPLANTABLE DEVICES (BIOPROSTHETIC HEART VALVES AND ANNULOPLASTY RINGS), RATHER THAN A TRUE POST-MARKET SURVEILLANCE REGISTRY. THROUGH THE REGISTRY, EDWARDS IS NOTIFIED WHEN THESE DEVICES ARE IMPLANTED. IN ADDITION, PATIENT AND/OR DEVICE STATUS MAY BE REPORTED TO THE REGISTRY VIA THE IMPLANTATION DATA CARDS. THE INFORMATION IS RECEIVED FROM VARIOUS SOURCES (E.G. SURGEON, HOSPITAL, AND PATIENT FAMILY MEMBERS) AND IS NOT RECEIVED IN THE FORM OF A CONVENTIONAL "CUSTOMER COMPLAINT". THE INFORMATION REPORTED MAY OR MAY NOT BE RELATED TO THE EDWARDS DEVICE. IN THIS CASE, IT WAS REPORTED THAT THE VALVE WAS EXPLANTED AFTER AN IMPLANT DURATION OF 10 YEARS DUE TO PROSTHETIC AORTIC VALVE STENOSIS, SECONDARY TO CALCIFICATION. THE HEALTHCARE PROVIDER ALSO NOTED THAT THIS EVENT WAS DUE TO AGE RELATED CALCIFICATION AND NOT A DEVICE MALFUNCTION. THE EXPLANTED DEVICE WAS REPLACED WITH A NEW EDWARDS BIOPROSTHETIC VALVE. NO OTHER DETAILS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 67613 | CARPENTIER-EDWARDS PERIMOUNT RSR PERICARDIAL BIOPROSTHESIS | REPLACEMENT HEART VALVE | DYE | EDWARDS LIFESCIENCES | 2800 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 67 YR | Hospitalization| R |