FDA Adverse Event Injury Summary report: N

CONSULTA CRT-D

MDR report key: 2964415 · Received February 15, 2013

Report

Report Number
3004209178-2013-02328
Event Type
Injury
Date Received
February 15, 2013
Report Date
December 20, 2012
Manufacturer
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
Product Code
NIK
PMA / PMN Number
P010031
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCT: 1488T COMPETITOR IMPLANTABLE PACING LEAD (B)(6) 2002; 6935 IMPLANTABLE TACHY LEAD (B)(6) 2010. (B)(4).

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. THE DEVICE MET 88% OF EXPECTED LONGEVITY. WITHOUT THE HISTORY OF THE PROGRAMMED SETTINGS THROUGHOUT ITS SERVICE LIFE, THERE IS NO WAY TO DETERMINE WHY THE LONGEVITY DID NOT MATCH THE PREDICTED MODEL.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THE DEVICE REACHED ELECTIVE REPLACEMENT INDICATOR (ERI) IN LESS THAN THREE YEARS AND THERE WAS POSSIBLE EARLY BATTERY DEPLETION. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
68333 CONSULTA CRT-D DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO NIK MEDTRONIC MED REL MEDTRONIC PUERTO RICO D224TRK

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R 4195 IMPLANTABLE PACING LEAD