FDA Adverse Event Malfunction Summary report: N

ORTHO SUMMIT SAMPLE HANDLING SYSTEM

MDR report key: 2964315 · Received February 15, 2013

Report

Report Number
2250051-2013-00032
Event Type
Malfunction
Date Received
February 15, 2013
Date of Event
January 17, 2013
Report Date
February 15, 2013
Manufacturer
HAMILTON BONADUZ AG
Product Code
MMH
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ON (B)(4) 2013 AN OCD FIELD ENGINEER (FE) ARRIVED AT THE CUSTOMER SITE AND NOTICED 2 TIP CLAMPS WERE BROKEN. THE FE REPLACED THE BROKEN TIP CLAMPS. THE FE PERFORMED ALL THE REQUIRED ADJUSTMENTS AND RAN SPLLD TESTS AND BOOT RUN TESTS WITHOUT ERROR. OPERATIONS VERIFIED AND ACCEPTED. QC PERFORMED BY CUSTOMER. REPAIRS HAVE RETURNED THIS INSTRUMENT TO EXPECTED OPERATION.

Description of Event or Problem · 1

THE ORTHO SUMMIT SAMPLE HANDLING SYSTEM INSTRUMENT DID NOT PIPETTE THE CORRECT AMOUNT OF SAMPLE AND DID NOT GENERATE AN ERROR MESSAGE. NO DEATH OR SERIOUS INJURY WAS ASSOCIATED WITH THIS INCIDENT. (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
67544 ORTHO SUMMIT SAMPLE HANDLING SYSTEM SAMPLE PIPETTOR MMH HAMILTON BONADUZ AG 936400

Patients

Seq Age Sex Outcome Treatment
1