ORTHO SUMMIT SAMPLE HANDLING SYSTEM
MDR report key: 2964315
·
Received February 15, 2013
Report
- Report Number
- 2250051-2013-00032
- Event Type
- Malfunction
- Date Received
- February 15, 2013
- Date of Event
- January 17, 2013
- Report Date
- February 15, 2013
- Manufacturer
- HAMILTON BONADUZ AG
- Product Code
- MMH
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
ON (B)(4) 2013 AN OCD FIELD ENGINEER (FE) ARRIVED AT THE CUSTOMER SITE AND NOTICED 2 TIP CLAMPS WERE BROKEN. THE FE REPLACED THE BROKEN TIP CLAMPS. THE FE PERFORMED ALL THE REQUIRED ADJUSTMENTS AND RAN SPLLD TESTS AND BOOT RUN TESTS WITHOUT ERROR. OPERATIONS VERIFIED AND ACCEPTED. QC PERFORMED BY CUSTOMER. REPAIRS HAVE RETURNED THIS INSTRUMENT TO EXPECTED OPERATION.
Description of Event or Problem · 1
THE ORTHO SUMMIT SAMPLE HANDLING SYSTEM INSTRUMENT DID NOT PIPETTE THE CORRECT AMOUNT OF SAMPLE AND DID NOT GENERATE AN ERROR MESSAGE. NO DEATH OR SERIOUS INJURY WAS ASSOCIATED WITH THIS INCIDENT. (B)(4).