PROGRAMMING SOFTWARE
Report
- Report Number
- 1644487-2013-00419
- Event Type
- Malfunction
- Date Received
- February 15, 2013
- Date of Event
- January 17, 2013
- Report Date
- January 17, 2013
- Manufacturer
- CYBERONICS
- Product Code
- LYJ
- PMA / PMN Number
- P970003
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
DEVICE FAILURE OCCURRED, BUT DID NOT CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY.
A FORMER COMPANY REPRESENTATIVE'S PROGRAMMING SYSTEM WAS RETURNED TO THE MANUFACTURER WHEN THEY LEFT THE COMPANY. PRODUCT ANALYSIS WAS COMPLETED ON THE WAND, HANDHELD AND FLASHCARD. NO VISUAL OR MECHANICAL ANOMALY WAS IDENTIFIED WITH THE WAND. CONTINUITY TESTING OF THE SERIAL DATA CABLE AND THE BATTERY CABLE PASSED. THE PROGRAMMING WAND PERFORMED ACCORDING TO FUNCTIONAL SPECIFICATIONS. DURING THE ANALYSIS IT WAS IDENTIFIED THAT THE SYNC CABLE CONNECTOR ON THE BOTTOM OF THE HANDHELD WAS DAMAGED. BECAUSE OF THE DAMAGE, THE HANDHELD WAS UNABLE TO RECEIVE POWER FROM THE AC ADAPTER. THE CAUSE FOR THE DAMAGE TO THE CONNECTOR IS MOST LIKELY ASSOCIATED WITH MISHANDLING OF THE DEVICE AS IT APPEARS THE CONNECTOR DETENTS ARE NOT BEING COMPRESSED WHEN REMOVING OR MANIPULATING THE SERIAL DATA CABLE, THUS PLACING AN EXTENSIVE AMOUNT OF FORCE ON THE PCB AND ATTACHED CABLE RECEPTACLE. NO OTHER ANOMALIES WERE IDENTIFIED. NO ANOMALIES ASSOCIATED WITH FLASHCARD SOFTWARE OR DATABASES WERE IDENTIFIED DURING THE FLASHCARD ANALYSIS. THE FLASHCARD AND SOFTWARE PERFORMED ACCORDING TO FUNCTIONAL SPECIFICATIONS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 66814 | PROGRAMMING SOFTWARE | PROGRAMMING COMPUTER | LYJ | CYBERONICS | MODEL 250 | 840143 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |