FDA Adverse Event Malfunction Summary report: N

MEGA SUTURECUT NEEDLE DRIVERINSTRUMENT

MDR report key: 2964027 · Received February 15, 2013

Report

Report Number
2955842-2013-00517
Event Type
Malfunction
Date Received
February 15, 2013
Date of Event
January 25, 2013
Report Date
January 25, 2013
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. ENGINEERING EVALUATION CONFIRMED THE ALLEGED COMPLAINT OF A WIRE STICKING OUT. ONE GRIP CLOSE CABLE WAS FOUND TO BE BROKEN AT THE DISTAL IDLERS. THE IDLER PULLEY WAS OBSERVED TO SPIN FREELY AND WAS NOT DAMAGED. A CABLE SEGMENT WAS FOUND TO BE STICKING OUT AT WRIST. OTHER CABLES AT THE INSTRUMENT'S WRIST WERE NOT DAMAGED. ENGINEERING EVALUATION ALSO FOUND SCRATCHES ON THE SURFACE OF THE DISTAL PULLEY. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER THE REPORTED BROKEN CABLE ISSUE IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DA VINCI SI TOTAL BENIGN HYSTERECTOMY PROCEDURE, THE SURGICAL STAFF ALLEGED THAT A WIRE WAS OBSERVED TO BE STICKING OUT FROM THE MEGA SUTURECUT NEEDLE DRIVER INSTRUMENT . THERE WERE NO MISSING OR FALLEN PIECES REPORTED. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED WITH A LARGE NEEDLE DRIVER INSTRUMENT AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
67164 MEGA SUTURECUT NEEDLE DRIVERINSTRUMENT ENDOSCOPIC INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420309-01 M10120829 647

Patients

Seq Age Sex Outcome Treatment
1 DA VINCI SI SURGICAL SYSTEM AND ACCESSORIES