FDA Adverse Event Summary report: N

RELIANCE 120 CART WASHER

MDR report key: 2963924 · Received February 15, 2013

Report

Report Number
9680353-2013-00014
Date Received
February 15, 2013
Date of Event
January 16, 2013
Report Date
February 15, 2013
Manufacturer
STERIS CANADA CORPORATION
Product Code
LDS
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A STERIS SERVICE TECHNICIAN ARRIVED ON SITE AND FOUND THAT THE WATER LEVEL SENSORS WERE NOT WORKING PROPERLY. THE SERVICE TECHNICIAN CLEANED AND REBUILT THE WATER FILL (SOLENOID) VALVE AND SUMP FLOAT (CHAMBER WATER LEVEL CONTROL). A TEST CYCLE WAS RUN AND THE UNIT WAS OPERATIONAL.

Description of Event or Problem · 1

THE USER FACILITY REPORTED THEIR RELIANCE CART WASHER WAS LEAKING WATER FROM THE BOTTOM OUT ONTO THE FLOOR. NO PROCEDURAL DELAYS OR CANCELLATIONS REPORTED. NO INJURY TO HOSPITAL STAFF OR PATIENTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
67437 RELIANCE 120 CART WASHER WASHER LDS STERIS CANADA CORPORATION