FDA Adverse Event
Summary report: N
RELIANCE 120 CART WASHER
MDR report key: 2963924
·
Received February 15, 2013
Report
- Report Number
- 9680353-2013-00014
- Date Received
- February 15, 2013
- Date of Event
- January 16, 2013
- Report Date
- February 15, 2013
- Manufacturer
- STERIS CANADA CORPORATION
- Product Code
- LDS
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
A STERIS SERVICE TECHNICIAN ARRIVED ON SITE AND FOUND THAT THE WATER LEVEL SENSORS WERE NOT WORKING PROPERLY. THE SERVICE TECHNICIAN CLEANED AND REBUILT THE WATER FILL (SOLENOID) VALVE AND SUMP FLOAT (CHAMBER WATER LEVEL CONTROL). A TEST CYCLE WAS RUN AND THE UNIT WAS OPERATIONAL.
Description of Event or Problem · 1
THE USER FACILITY REPORTED THEIR RELIANCE CART WASHER WAS LEAKING WATER FROM THE BOTTOM OUT ONTO THE FLOOR. NO PROCEDURAL DELAYS OR CANCELLATIONS REPORTED. NO INJURY TO HOSPITAL STAFF OR PATIENTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 67437 | RELIANCE 120 CART WASHER | WASHER | LDS | STERIS CANADA CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |