FDA Adverse Event Malfunction Summary report: N

INFUSOR

MDR report key: 2963652 · Received February 15, 2013

Report

Report Number
1416980-2013-03923
Event Type
Malfunction
Date Received
February 15, 2013
Date of Event
January 1, 2013
Report Date
January 24, 2013
Manufacturer
BAXTER HEALTHCARE - IRVINE
Product Code
MEB
PMA / PMN Number
K071222
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE SAMPLE WAS NOT RETURNED TO BAXTER FOR EVALUATION. THEREFORE, THE REPORTED CONDITION COULD NOT BE CONFIRMED AND THE ROOT CAUSE COULD NOT BE DETERMINED.

Additional Manufacturer Narrative · 1

(B)(4). THE BATCH REVIEW REVEALED THAT ALL OF THE ACCEPTANCE CRITERIA WERE MET TO RELEASE THE LOT. THERE WERE NO NON-CONFORMANCES, FAILURES, REWORK, OR DEVIATIONS RELATED TO THE LOT. THERE WERE NO CHANGES TO SPECIFICATIONS, TEST METHODS, PROCESS, EQUIPMENT, OR RAW MATERIALS THAT COULD BE ASSOCIATED WITH THE REPORTED CONDITION. A REQUEST FOR THE RETURN OF THE DEVICE HAS BEEN MADE. SHOULD THE DEVICE BE RECEIVED BY BAXTER FOR EVALUATION, A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF AN EVALUATION OR IF ANY ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 1

BAXTER (B)(4) RECEIVED A REPORT THAT AN INFUSOR OVERINFUSED DURING PATIENT USE. THE TOTAL FILL VOLUME OF 125 ML WAS INFUSED OVER THE COURSE OF 9 HOURS RATHER THAN 12 HOURS. THE CONCOMITANT MEDICAL PRODUCTS ARE CURRENTLY UNKNOWN. THERE WAS PATIENT INVOLVEMENT, BUT THERE WAS NO REPORT OF PATIENT INJURY, MEDICAL INTERVENTION NECESSARY, OR ADVERSE REACTION IN ASSOCIATION WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
67137 INFUSOR PUMP, INFUSION, ELASTOMERIC MEB BAXTER HEALTHCARE - IRVINE 12E027

Patients

Seq Age Sex Outcome Treatment
1