MDR report key: 2963642 · Received January 2, 2013

Report

Report Number
1219977-2012-00132
Event Type
Injury
Date Received
January 2, 2013
Date of Event
February 1, 2008
Report Date
June 29, 2008
Manufacturer
ATRIUM MEDICAL CORP.
Product Code
DSY
PMA / PMN Number
K992960
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

A SMALL SEGMENT OF THE GRAFT (ABOUT 5CM LONG) WAS RECEIVED BY ATRIUM TO DETERMINE THE CAUSE OF PERFORATIONS IN THE GRAFT WALL AND TO DETERMINE IF THE GRAFT WALL WAS THIN AT THE AREA IN CONTACT WITH THE REPORTED PSEUDOANEURYSM. THE SAMPLE WAS EXAMINED GROSSLY PRIOR TO BEING EXAMINED MICROSCOPICALLY. ATRIUM RECEIVED THE SAMPLE COVERED WITH TISSUE GROWN ON THE OUTER SURFACE OF THE GRAFT. THE WALL THICKNESS OF THE EXPLANTED GRAFT SAMPLE FELL WITHIN SPECIFICATION. THIS INDICATES THAT THE GRAFT HAD NO WALL DEFECT IN TERMS OF ONE SEGMENT OF THE WALL BEING TOO THIN AS INDICATED. THERE WERE NO GRAFT DEFECTS FOUND THAT COULD BE DETECTED USING EITHER THE DIGITAL IMAGING MICROSCOPE OR GROSS VISUAL EXAMINATION, OTHER THAN TWO PERFORATIONS THAT ARE CONSISTENT WITH DIALYSIS NEEDLE STICKS (SIZE OF HOLES MEASURED TO BE 0.013" [0.33MM]). THE LOT HISTORY OF THE GRAFT WAS REVIEWED AND THE PRODUCT WAS FOUND TO HAVE MET ALL SPECIFICATIONS. LITERATURE REVIEW SUGGESTS THAT PSEUDOANEURYSM OCCURRENCE IS NOT UNUSUAL AND HAS MANY CAUSES THAT ARE OFTEN IATROGENIC OR BIOLOGIC IN ORIGIN.

Description of Event or Problem · 1

THE ADVANTA VXT WITH GDS WAS IMPLANTED IN THE LEFT BRACHIAL REGION OF A (B)(6) FEMALE PT WHO WAS TAKING ORAL STEROIDS, FOR USE AS AN A-V SHUNT, APPROX 20CM IN LENGTH, ON (B)(6) 2007. A POTENTIAL GRAFT ANEURYSM WAS NOTICED IN (B)(6) 2008 AND THE PT WAS FOLLOWED WITHOUT TREATMENT. AN ANGIOGRAM WAS PERFORMED IN (B)(6) 2008 DUE TO AN INCREASE IN SIZE OF THE ANEURYSM, NOW APPROX 2CM. SURGICAL REVISION WAS PERFORMED ON (B)(6) AT WHICH TIME TWO PERFORATIONS IN THE GRAFT WALL WERE SEEN. THE OPERATOR COMMENTED THAT THE GRAFT WALL LOOKED THIN IN THE AREA OF THE PSEUDOANEURYSM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1644 ADVANTA VXT, GDS DSY ATRIUM MEDICAL CORP. 22017 AG05211

Patients

Seq Age Sex Outcome Treatment
1 39 YR Required Intervention ORAL STEROIDS