HOMECHOICE PRO
Report
- Report Number
- 1416980-2013-03914
- Event Type
- Malfunction
- Date Received
- February 15, 2013
- Date of Event
- January 9, 2013
- Report Date
- January 31, 2013
- Manufacturer
- BAXTER HEALTHCARE - LARGO
- Product Code
- FKX
- PMA / PMN Number
- K102936
- Removal / Correction Number
- 1423500-01/08/10-001-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PL
- Reporter Occupation
- PATIENT
Narratives
(B)(4). EVALUATION: THE DEVICE WAS RETURNED TO BAXTER, AND THE EVALUATION IS COMPLETE. THIS COMPLAINT IS AN ANCILLARY SERVICE EVENT. THE REPORTED DIFFICULTY OF AN IIPV EVENT WAS CONFIRMED THROUGH AN EVENT HISTORY LOG REVIEW. THE CAUSE WAS USE ERROR, TIDAL TOTAL UF REMOVAL SET TOO LOW. IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW-UP WILL BE SENT.
IT WAS REPORTED THAT DURING SERVICING OF A RETURNED HOMECHOICE MACHINE, ONE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED WHICH OCCURRED IN THE THERAPY INITIATED ON (B)(6) 2013, 21:57:37. DURING NIGHT DRAIN CYCLE SIX, THE PATIENT'S ULTRAFILTRATION READING WAS 2107ML, INDICATING THE HOME PATIENT (HP) DRAINED 1707ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 2000ML. THIS INFORMATION MEETS IIPV CRITERIA. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 67125 | HOMECHOICE PRO | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - LARGO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |