PWRD ECH FLEX 60MM
Report
- Report Number
- 3005075853-2013-00636
- Event Type
- Malfunction
- Date Received
- February 15, 2013
- Date of Event
- January 21, 2013
- Report Date
- January 21, 2013
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GDW
- PMA / PMN Number
- K110385
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- OTHER
Narratives
(B)(4).
(B)(4). THE ANALYSIS FOUND THAT ONE DEVICE WAS RETURNED IN GOOD VISUAL CONDITION AND WITH NO CARTRIDGE RELOAD LOADED ON THE DEVICE. THE DEVICE WAS TESTED FOR FUNCTIONALITY IN THE STRAIGHT POSITION WITH A TEST CARTRIDGE RELOAD AND ACHIEVED ITS COMPLETE FIRING SEQUENCE WITHOUT ANY DIFFICULTIES. THE STAPLE LINE AND CUT LINE WERE COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORM SHAPE. EVENT COULD NOT BE CONFIRMED AS THE DEVICE PERFORMED WITHOUT ANY ANOMALIES NOTED DURING FUNCTIONAL TESTING. THIS REPORT IS NOT INTENDED TO DENY THAT YOU EXPERIENCED A PROBLEM WITH THE DEVICE. THERE MAY HAVE BEEN OTHER CIRCUMSTANCES OR ISSUES THAT OCCURRED DURING THE USE OF THE DEVICE THAT WE WERE UNABLE TO DUPLICATE DURING OUR LABORATORY ANALYSIS. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, ON THE THIRD FIRING THE INSTRUMENT DE-ARTICULATED WHEN THE SAFETY CATCH WAS RELEASED AND THE POWER BUTTON PUSHED (DURING FIRING). THE SURGEON REVERSED THE FIRING, AND COMPLETED THE PROCEDURE WITH A NEW STAPLER AND A NEW RELOAD. NO DELAY TO SURGERY. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 66957 | PWRD ECH FLEX 60MM | STAPLE, IMPLANTABLE | GDW | ETHICON ENDO-SURGERY, LLC. | NA | J4CMOW |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |