PRISM HTLV-I/HTLV-II
Report
- Report Number
- 1415939-2013-00053
- Event Type
- Malfunction
- Date Received
- February 15, 2013
- Report Date
- February 5, 2013
- Manufacturer
- ABBOTT LABORATORIES
- Product Code
- MTP
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS
- Reporter Occupation
- OTHER
Narratives
A RETAINED KIT OF PRISM HTLV-1 / HTLV-2 REAGENT, LIST NUMBER 06A53-48 LOT NUMBER 21105LI00 WAS CALIBRATED ON ONE PRISM ANALYZER. TO EVALUATE THE PERFORMANCE WITH REGARD TO SPECIFICITY, REPLICATES OF AN HTLV-1 / HTLV-2 SPECIFICITY PANEL WERE TESTED. RESULTS WERE WITHIN TYPICAL RANGES AND NO FALSE REACTIVE RESULTS WERE OBTAINED. A REVIEW OF THE HUMAN NEGATIVE POPULATION, ROUTINELY EVALUATED DURING RELEASE TESTING, REVEALED NO INDICATIONS THAT THE SPECIFICITY OF THE REAGENT LOT MIGHT BE ADVERSELY AFFECTED. THE MEAN VALUE ACROSS ALL PANEL MEMBERS FOR THIS LOT NUMBER CORRESPONDS TO THE MEAN OBSERVED FOR ALL LOT NUMBERS. A REVIEW OF COMPLAINT TRACKING AND TRENDING METRICS WAS PERFORMED AND IDENTIFIED NO ADVERSE TRENDS IN CONJUNCTION WITH THE COMPLAINT ISSUE CURRENTLY UNDER EVALUATION. ADDITIONALLY, THE PERFORMANCE WAS INVESTIGATED BY COMPLETING A REVIEW OF MANUFACTURING AND TESTING RECORDS FOR THE AFFECTED REAGENT LOT. THIS REVIEW DID NOT REVEAL ANY EVENTS OR ISSUES THAT MAY HAVE IMPACTED REAGENT PERFORMANCE IN RELATION TO THE CURRENT COMPLAINT ISSUE. THE PRISM HTLV-1/HTLV-2 ASSAY PACKAGE INSERT CONTAINS INFORMATION TO ADDRESS THE CURRENT CUSTOMER ISSUE. BASED ON THE RESULTS OF THE CURRENT EVALUATION, IT HAS BEEN DETERMINED THAT THE PRISM HTLV-1 / HTLV-2 REAGENT, LIST NUMBER 06A53-48, LOT NUMBER 21105LI00, IS PERFORMING ACCEPTABLY.
AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. THIS REPORT IS BEING FILED ON AN INTERNATIONAL PRODUCT, LIST NUMBER 06A53 THAT HAS A SIMILAR PRODUCT DISTRIBUTED IN THE US, LIST NUMBER 06E50. (B)(4).
THE CUSTOMER REPORTS AN INCREASE IN REPEAT (B)(6) RATES FOR THE PRISM HTLV-I/HTLV-II ASSAY WHEN THEY SWITCHED TO REAGENT LOT 21105LI00. THIS OCCURRED DURING THE TIMEFRAME OF (B)(6) 2013. THIS HAS RESULTED IN DONORS WHO PREVIOUSLY TESTED (B)(6) TO NOW TEST (B)(6). THE CUSTOMER PROVIDED THE FOLLOWING EXAMPLE: DONOR 7: INITIAL (B)(6) RESULT OF (B)(6); REPEAT (B)(6) RESULTS OF (B)(6). PREVIOUS (B)(6) RESULT OF (B)(6) ON (B)(6) 2011 WITH REAGENT LOT 89684HN00. TO DATE, THE CUSTOMER HAS SEEN THEIR REPEAT (B)(6). THERE IS NO FURTHER IMPACT TO DONOR MANAGEMENT REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 68973 | PRISM HTLV-I/HTLV-II | MTP | ABBOTT LABORATORIES | 21105LI00 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | PRISM 6 CHANNEL CE LN: 06A36-04 SN: (B)(4)| PRISM 6 CHANNEL CE LN: 06A36-04 SN: (B)(4) |