FDA Adverse Event Malfunction Summary report: N

ARCHITECT CA 19-9XR

MDR report key: 2963115 · Received February 15, 2013

Report

Report Number
1415939-2013-00050
Event Type
Malfunction
Date Received
February 15, 2013
Report Date
February 6, 2013
Manufacturer
ABBOTT LABORATORIES
Product Code
NIG
PMA / PMN Number
K052000
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

EVALUATION OF THE CUSTOMER ISSUE INCLUDED A REVIEW OF THE COMPLAINT TEXT, A SEARCH FOR SIMILAR COMPLAINTS, SPECIFICITY TESTING, AND A REVIEW OF LABELING. A REVIEW OF CUSTOMER COMPLAINTS RECEIVED TO-DATE DID NOT IDENTIFY AN INCREASE IN COMPLAINT ACTIVITY RELATED TO DISCREPANT PATIENT RESULTS WITH ARCHITECT CA19-9XR, LIST 2K91-25, LOT 14780M500. ACCURACY TESTING WAS COMPLETED USING AN INTERNAL CA19-9XR PANEL AND RETAINED KITS OF REAGENT LOT 14780M500, WHICH MET ACCEPTANCE CRITERIA. FURTHER INVESTIGATION IS NOT REQUIRED. BASED ON THE RESULTS OF THIS EVALUATION, THE ARCHITECT CA19-9XR ASSAY, LIST 2K91-25, LOT 14780M500, IS PERFORMING AS INTENDED AND NO PRODUCT DEFICIENCY OR MALFUNCTION WAS IDENTIFIED.

Additional Manufacturer Narrative · 1

(B)(4). A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. (B)(6). THIS REPORT IS BEING FILED ON AN INTERNATIONAL PRODUCT, LIST NUMBER 2K91-25, THAT HAS A SIMILAR PRODUCT DISTRIBUTED IN THE US, LIST NUMBER 2K91-27.

Description of Event or Problem · 1

A FALSELY ELEVATED ARCHITECT CA19-9 RESULT OF 70 U/ML WAS GENERATED ON ONE PATIENT SAMPLE DURING A HEALTH SCREENING. UPON EXAMINATION AND REPEAT TESTING OF THE PATIENT THE ARCHITECT CA19-9 RESULT GENERATED LESS THAN 2.00U/ML. THE PATIENT'S CRP RESULT WAS NORMAL. THERE WAS NO REPORTED IMPACT TO PATIENT MANAGEMENT. THERE WAS NO ADDITIONAL PATIENT INFORMATION PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
68438 ARCHITECT CA 19-9XR NIG ABBOTT LABORATORIES 14780M500

Patients

Seq Age Sex Outcome Treatment
1 SERIAL # (B)(4)| SERIAL # (B)(4)| ARCHITECT I1000SR ANALYZER, LN 01L86-01| ARCHITECT I1000SR ANALYZER, LN 01L86-01