HOMECHOICE AUTOMATED PD SET WITH CASSETTE
Report
- Report Number
- 1416980-2013-03909
- Event Type
- Malfunction
- Date Received
- February 15, 2013
- Date of Event
- January 31, 2013
- Report Date
- January 31, 2013
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- FKX
- PMA / PMN Number
- K923065
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- PATIENT
Narratives
(B)(4). THE PROBLEM WAS CONFIRMED. THE ROOT CAUSE WAS UNDETERMINED. PER THE CUSTOMER THE SAMPLE WAS DISCARDED AND THE LOT NUMBER WAS UNKNOWN, THEREFORE NO EVALUATION OR BATCH REVIEW COULD BE PERFORMED. A FOLLOW-UP REPORT WILL BE FILED IF ANY ADDITIONAL INFORMATION BECOMES AVAILABLE.
THE CUSTOMER CONTACTED BAXTER'S TECHNICAL SERVICE CENTER REGARDING A SYSTEM ERROR (SE) 2240 ALARM, WHICH OCCURRED ON THE HOMECHOICE (HC) DURING USE, DURING DWELL 3 OF 5. THE CAREGIVER (CG) STATED A SUPPLY BAG CAME LOOSE. THE CG STATED THEY WOULD NOT RESTART THERAPY THAT NIGHT. THE BAXTER TECHNICAL SERVICE REPRESENTATIVE (TSR) ASSISTED THE CG TO END THERAPY AND THE TSR ADVISED THE CG TO DISPOSE OF ALL CURRENT SUPPLIES USED. THE PATIENT WAS CONNECTED EITHER AT THE TIME OF THE ALARM OR OBSERVED AIR. THE PATIENT LINE WAS PROPERLY PRIMED BEFORE CONNECTING. PATIENT EXTENSION LINES WERE NOT USED. THE PATIENT DID NOT DISCONNECT PRIOR TO THE ALARM OR OBSERVED AIR. THE SUPPLIES WERE NOT DAMAGED BY AN OUTLET PORT CLAMP OR AN ASSIST DEVICE USED TO MAKE THE CONNECTIONS. PROPER PROCEDURES PER THE USER MANUAL WERE REVIEWED WITH THE REPORTER. IT WAS UNKNOWN HOW THE HP WOULD COMPLETE THERAPY. THE SOLUTION TO THIS ISSUE WAS PROVIDED OVER THE PHONE AND A SWAP OF THE DEVICE WAS NOT NECESSARY. THERE WAS PATIENT INVOLVEMENT BUT NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED AT THE TIME OF THE INITIAL REPORT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 66825 | HOMECHOICE AUTOMATED PD SET WITH CASSETTE | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - MOUNTAIN HOME |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 77 YR | HOMECHOICE |