FDA Adverse Event Injury Summary report: N

MENISCAL CINCH, CURVED TIP

MDR report key: 2962944 · Received February 15, 2013

Report

Report Number
1220246-2013-00011
Event Type
Injury
Date Received
February 15, 2013
Date of Event
January 17, 2013
Report Date
January 24, 2013
Manufacturer
ARTHREX, INC.
Product Code
GAT
PMA / PMN Number
K073149
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

PATIENT DEMOGRAPHICS (AGE AT TIME OF EVENT, DATE OF BIRTH, GENDER, WEIGHT) WERE REQUESTED BUT NOT PROVIDED. NO FURTHER PATIENT INFORMATION WAS PROVIDED AT THE TIME OF THIS REPORT OR MADE AVAILABLE IN RESPONSE TO FOLLOW-UP COMMUNICATION. NO ADDITIONAL ADVERSE CONSEQUENCES HAVE BEEN REPORTED FROM THIS EVENT. THIS DEVICE IS USED FOR TREATMENT. THE DEVICE WAS REQUESTED FOR EVALUATION BUT WAS NOT RETURNED, THEREFORE THE COMPLAINANT'S EVENT COULD NOT BE VERIFIED. THE CAUSE OF THE EVENT COULD NOT BE DETERMINED FROM THE INFORMATION AVAILABLE AND WITHOUT DEVICE EVALUATION. DEVICE HISTORY RECORD REVIEW REVEALED NOTHING RELEVANT TO THIS EVENT. THE MOST LIKELY CAUSE(S) OF THIS TYPE OF EVENT INCLUDE INADVERTENT FRAYING, NICKING OR CUTTING OF THE SUTURE WHILE ADVANCING THE INSERTION SPEARS OR EXCESSIVE FORCE TO THE SUTURE DURING TENSIONING OR SUTURE SLACK REMOVAL. THIS IS THE FIRST COMPLAINT OF THIS TYPE FOR THIS PART/LOT COMBINATION. THE POTENTIAL CAUSE(S) OF THIS EVENT WILL BE COMMUNICATED TO THE EVENT REPORTER. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A FOLLOW-UP REPORT WILL BE SUBMITTED. DISCARDED BY THE FACILITY.

Description of Event or Problem · 1

IT WAS REPORTED THAT BOTH IMPLANTS OF THE MENISCAL CINCH WERE SUCCESSFULLY DEPLOYED INTO MENISCUS. WHEN PULLING THE SUTURE TO TIGHTEN, IT BECAME DETACHED FROM ONE IMPLANT WHICH WAS LEFT UN-RETRIEVED BEHIND MENISCUS. ANOTHER MENISCAL CINCH USED TO FINISH SURGERY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
66744 MENISCAL CINCH, CURVED TIP SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE GAT ARTHREX, INC. 433638

Patients

Seq Age Sex Outcome Treatment
1 Other