FDA Adverse Event Malfunction Summary report: N

XENIUM XPM

MDR report key: 2962848 · Received February 14, 2013

Report

Report Number
1416980-2013-03892
Event Type
Malfunction
Date Received
February 14, 2013
Date of Event
January 29, 2013
Report Date
January 29, 2013
Manufacturer
NIPRO CORPORATION *USD*
Product Code
KDI
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BR
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THIS COMPLAINT WAS NOT CONFIRMED AND THE ROOT CAUSE COULD NOT BE DETERMINED. THE SAMPLE WAS NOT RETURNED TO BAXTER; THEREFORE NO EVALUATION COULD BE PERFORMED. HOWEVER, A RETAINED SAMPLE FROM THE SAME LOT WAS VISUALLY AND FUNCTIONALLY TESTED BY NIPRO WITH NO DAMAGE OR LEAKS NOTED. A MANUFACTURING BATCH RECORD REVIEW WAS PERFORMED BY NIPRO WITH NO ISSUES OR ABNORMALITIES NOTED DURING THE MANUFACTURE OF THIS LOT. THIS ISSUE IS BEING INVESTIGATED THROUGH CAPA.

Additional Manufacturer Narrative · 1

(B)(4). THE ACTUAL SAMPLE WAS NOT AVAILABLE; HOWEVER, A COMPANION SAMPLE HAS BEEN REPORTED TO BE AVAILABLE AND REQUESTED FOR EVALUATION. THE COMPANION SAMPLE HAS NOT BEEN RECEIVED FOR EVALUATION AS OF YET. A FOLLOW-UP REPORT WILL BE FILED IF ADDITIONAL INFORMATION IS RECEIVED. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PATIENT HAD A BLOOD LEAK DURING HEMODIALYSIS THERAPY, THE LEAK WAS OBSERVED ON THE SEALING OF THE VENOUS CAP. THE ESTIMATED BLOOD LOSS FOR THE PATIENT WAS 50ML. THE TREATMENT WAS RESUMED WITH A NEW DIALYZER FROM A DIFFERENT BATCH. THE NURSE INFORMED THAT THE ISSUE WAS OBSERVED DURING THE DIALYZERS FIRST USE, AND THE DIALYZERS WERE NOT REUSED. THE NURSE INFORMED THAT THE DIALYZER PASSED THE LEAK TEST PRIOR TO PATIENT USE. THERE WAS PATIENT INVOLVEMENT; HOWEVER, NO PATIENT MEDICAL INJURY OR ADVERSE EVENT WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
64486 XENIUM XPM DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM KDI NIPRO CORPORATION *USD* 12G19B

Patients

Seq Age Sex Outcome Treatment
1