FDA Adverse Event Malfunction Summary report: N

RIATA ST ACTIVE FIXATION

MDR report key: 2962819 · Received February 14, 2013

Report

Report Number
2938836-2013-00111
Event Type
Malfunction
Date Received
February 14, 2013
Date of Event
January 24, 2013
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A FOLLOW UP, NOISE WAS NOTED. NOISE WAS REPRODUCED DURING POCKET MANIPULATION. DURING A LEAD REVISION INSULATION ABRASION NEAR THE CONNECTOR WAS OBSERVED. THE LEAD WAS CAPPED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
64480 RIATA ST ACTIVE FIXATION DEFIBRILLATION LEAD NVY ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 7002/65 NA

Patients

Seq Age Sex Outcome Treatment
1 73 YR