Summary report: N
RIATA ST ACTIVE FIXATION
MDR report key: 2962816
·
Received February 14, 2013
Report
- Report Number
- 2938836-2013-00106
- Event Type
- Malfunction
- Date Received
- February 14, 2013
- Date of Event
- September 20, 2012
- Manufacturer
- ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
- Product Code
- NVY
- PMA / PMN Number
- P950022
- Removal / Correction Number
- Z0458
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
A PARTIAL LEAD WITH THE CONNECTOR PIN MEASURING 13.9CM WAS RETURNED FOR ANALYSIS. EXTERNAL INSULATION ABRASION WAS FOUND AT 12.3-13.2CM FROM THE CONNECTOR PIN, CONSISTENT WITH LEAD FRICTION TO THE ICD CAN. THE ETFE COATING WAS ABRADED AT THIS LOCATION. EXTERNAL INSULATION ABRASION WERE FOUND AT 11.2-11.4CM AND 12.9-13.2CM FROM THE CONNECTOR PIN, CONSISTENT WITH FRICTION TO THE ICD CAN. THE ETFE COATING WAS INTACT AT THESE LOCATIONS.
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. DEVICE EVALUATION ANTICIPATED BUT NOT YET BEGUN.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE LEAD WAS CAPPED DUE TO EXTERNALIZED CONDUCTORS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 64479 | RIATA ST ACTIVE FIXATION | DEFIBRILLATION LEAD | NVY | ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION | 7000/60 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 51 YR |