FDA Adverse Event Malfunction Summary report: N

RIATA ST ACTIVE FIXATION

MDR report key: 2962816 · Received February 14, 2013

Report

Report Number
2938836-2013-00106
Event Type
Malfunction
Date Received
February 14, 2013
Date of Event
September 20, 2012
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
Z0458
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A PARTIAL LEAD WITH THE CONNECTOR PIN MEASURING 13.9CM WAS RETURNED FOR ANALYSIS. EXTERNAL INSULATION ABRASION WAS FOUND AT 12.3-13.2CM FROM THE CONNECTOR PIN, CONSISTENT WITH LEAD FRICTION TO THE ICD CAN. THE ETFE COATING WAS ABRADED AT THIS LOCATION. EXTERNAL INSULATION ABRASION WERE FOUND AT 11.2-11.4CM AND 12.9-13.2CM FROM THE CONNECTOR PIN, CONSISTENT WITH FRICTION TO THE ICD CAN. THE ETFE COATING WAS INTACT AT THESE LOCATIONS.

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. DEVICE EVALUATION ANTICIPATED BUT NOT YET BEGUN.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LEAD WAS CAPPED DUE TO EXTERNALIZED CONDUCTORS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
64479 RIATA ST ACTIVE FIXATION DEFIBRILLATION LEAD NVY ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 7000/60 NA

Patients

Seq Age Sex Outcome Treatment
1 51 YR