RIATA ST ACTIVE FIXATION

MDR report key: 2962797 · Received February 14, 2013

Report

Report Number
2938836-2013-00108
Event Type
Malfunction
Date Received
February 14, 2013
Date of Event
December 17, 2012
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
Z0458
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

EXTERNAL INSULATION ABRASIONS WERE FOUND AT 46.3 - 46.6CM, 47.3 - 47.6CM, AND 48.0 - 48.2CM FROM THE DISTAL TIP, CONSISTENT WITH LEAD FRICTION TO THE ICD CAN. THE ETFE COATING WAS INTACT AT THESE LOCATIONS.

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. DEVICE EVALUATION ANTICIPATED BUT NOT YET BEGUN.

Description of Event or Problem · 1

AN ASYMPTOMATIC PATIENT PRESENTED IN THE OR FOR DEVICE CHANGE-OUT DUE TO NORMAL ERI. EXTERNALIZED CONDUCTORS WERE NOTED. NO ELECTRICAL ANOMALIES WERE DETECTED. LEAD WAS EXPLANTED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
64716 RIATA ST ACTIVE FIXATION DEFIBRILLATION LEAD NVY ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 7001/65 NA

Patients

Seq Age Sex Outcome Treatment
1 80 YR